Page 1
S S e e c c u u r r e e ® ® C C o o n n n n e e c c t t ™ ™ O O p p e e r r a a t t i i o o n n s s / / M M a a i i n n t t e e n n a a n n c c e e M M a a n n u u a a l l 521200380100 Connected Hospital®...
Page 3
Serial number European medical device CE mark Authorized representative in the European Community For US Patents see www.stryker.com/patents Manufacturer Protection from liquid splash I I P P X X 4 4 Class II electrical equipment: equipment in which protection against electric shock does not rely...
Page 5
T T a a b b l l e e o o f f C C o o n n t t e e n n t t s s Symbols ................................1 Warning/Caution/Note Definition ........................2 Summary of safety precautions .......................2 Introduction ..............................3 Product description ..........................3 Expected service life..........................3...
Page 6
W W a a r r n n i i n n g g / / C C a a u u t t i i o o n n / / N N o o t t e e D D e e f f i i n n i i t t i i o o n n The words W W A A R R N N I I N N G G , C C A A U U T T I I O O N N , and N N O O T T E E carry special meanings and should be carefully reviewed.
Page 7
I I n n t t r r o o d d u u c c t t i i o o n n This manual assists you with the operation or maintenance of your Stryker product. Read this manual before operating or maintaining this product.
Page 8
S S e e c c u u r r e e C C o o n n n n e e c c t t : 18 VDC, 1.67A Uses infrared (IR) LED and Bluetooth based on Stryker proprietary communication scheme...
Page 9
Competent authority of the European Member State where the user and/or patient is established. To view your operations or maintenance manual online, see https://techweb.stryker.com/. Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication.
Page 10
C C A A U U T T I I O O N N - - Always match the dip-switches on SB1 and SB2 to the connected product configuration to avoid the risk of head wall damage. N N o o t t e e - - To confirm dip-switch configuration, contact Stryker customer service or technical support ( Contact information (page 5)).
Page 11
4 in. - 5 in. (10.2 cm - 12.7 cm) ± 0.5 in. (± 1.3 cm) F F i i g g u u r r e e 3 3 – – M M o o u u n n t t i i n n g g s s p p e e c c i i f f i i c c a a t t i i o o n n s s 6.
Page 12
I I n n s s t t a a l l l l a a t t i i o o n n c c h h e e c c k k l l i i s s t t Follow this checklist for the 521200380100 S S e e c c u u r r e e C C o o n n n n e e c c t t : Confirm that you do not have any unused components after installation.
Page 13
O O p p e e r r a a t t i i o o n n S S e e c c u u r r e e C C o o n n n n e e c c t t i i n n d d i i c c a a t t o o r r s s / / f f u u n n c c t t i i o o n n s s The S S e e c c u u r r e e C C o o n n n n e e c c t t has indicator symbols and a nurse call cord out cancel button (A) that is located on top of the product.
Page 14
N N o o t t e e - - Stryker service tool required if you use the S S e e c c u u r r e e C C o o n n n n e e c c t t scanner.
Page 15
N N o o t t e e - - Stryker service tool required if you use the S S e e c c u u r r e e C C o o n n n n e e c c t t scanner.
Page 16
Remove product from service before you perform the preventive maintenance inspection. Check all items listed during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks more often based on your level of product usage. Service only by qualified personnel.
Page 17
C C l l e e a a n n i i n n g g C C A A U U T T I I O O N N • Do not clean, disinfect, service, or perform maintenance while the product is in use. •...
Page 18
D D i i s s i i n n f f e e c c t t i i n n g g C C A A U U T T I I O O N N • Do not clean, disinfect, service, or perform maintenance while the product is in use. •...
Page 19
T T r r o o u u b b l l e e s s h h o o o o t t i i n n g g P P r r o o b b l l e e m m S S o o l l u u t t i i o o n n 1.
Page 20
S S e e r r v v i i c c e e R R e e p p l l a a c c i i n n g g t t h h e e b b a a t t t t e e r r y y ( ( 3 3 7 7 - - p p i i n n c c o o n n n n e e c c t t o o r r , , l l e e f f t t s s i i d d e e ) ) T T o o o o l l s s r r e e q q u u i i r r e e d d : : •...
Page 21
F F i i g g u u r r e e 8 8 – – R R e e p p l l a a c c i i n n g g t t h h e e b b a a t t t t e e r r y y ( ( l l e e f f t t ) ) 8.
Page 22
R R e e p p l l a a c c i i n n g g t t h h e e b b a a t t t t e e r r y y ( ( 3 3 7 7 - - p p i i n n c c o o n n n n e e c c t t o o r r , , r r i i g g h h t t s s i i d d e e ) ) T T o o o o l l s s r r e e q q u u i i r r e e d d : : •...
Page 23
S S e e c c u u r r e e C C o o n n n n e e c c t t l l o o c c a a t t o o r r a a s s s s e e m m b b l l y y , , l l e e f f t t 521200380105 Rev AC (Reference only) Torque item AB to 0.34 ±...
Page 24
Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 25
Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 26
I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 300900380940 Wallside infrared board 521200380111...
Page 27
I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 521200380803 37-pin cable assembly 521200380806...
Page 28
S S e e c c u u r r e e C C o o n n n n e e c c t t l l o o c c a a t t o o r r a a s s s s e e m m b b l l y y , , r r i i g g h h t t 521200380110 Rev AC (Reference only) Torque item AB Torque item AD...
Page 29
Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 30
Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 31
I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 300900380940 Wallside infrared board 521200380111...
Page 32
I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 521200380801 Battery cable assembly 521200380802...
Page 33
W W i i r r e e l l e e s s s s n n o o t t i i f f i i c c a a t t i i o o n n s s For product equipped with wireless communication technology, these statements apply to the countries as indicated: C C o o u u n n t t r r y y N N o o t t i i f f i i c c a a t t i i o o n n...
Page 34
E E M M C C i i n n f f o o r r m m a a t t i i o o n n W W A A R R N N I I N N G G •...
Page 35
G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ’ ’ s s d d e e c c l l a a r r a a t t i i o o n n - - e e l l e e c c t t r r o o m m a a g g n n e e t t i i c c i i m m m m u u n n i i t t y y Main power quality should ±0.5 kV, ±1 kV lines to lines ±0.5 kV, ±1 kV lines to lines...
Page 36
Portable and mobile RF communications equipment should follow the guidance in the table titled “Recommended separation distances between portable and mobile RF communication equipment and the 521200380100 S S e e c c u u r r e e C C o o n n n n e e c c t t locator.” If the mobile service is not listed in the table, the recommended separation...
R R e e c c o o m m m m e e n n d d e e d d s s e e p p a a r r a a t t i i o o n n d d i i s s t t a a n n c c e e s s b b e e t t w w e e e e n n p p o o r r t t a a b b l l e e a a n n d d m m o o b b i i l l e e R R F F c c o o m m m m u u n n i i c c a a t t i i o o n n e e q q u u i i p p m m e e n n t t a a n n d d t t h h e e 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 0 0 0 0 S S e e c c u u r r e e C C o o n n n n e e c c t t l l o o c c a a t t o o r r The 521200380100 S S e e c c u u r r e e C C o o n n n n e e c c t t locator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
Page 38
R R e e c c y y c c l l i i n n g g p p a a s s s s p p o o r r t t 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 0 0 5 5 / / 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 1 1 0 0 Rev AC R R e e c c y y c c l l a a b b l l e e p p a a r r t t I I m m p p o o r r t t a a n n t t...
7 7 0 0 0 0 0 0 0 0 0 0 8 8 8 8 6 6 7 7 9 9 5 5 Rev AB R R e e c c y y c c l l a a b b l l e e p p a a r r t t I I m m p p o o r r t t a a n n t t Q Q u u a a n n t t i i t t y y I I t t e e m m...
Page 40
7 7 0 0 0 0 0 0 0 0 0 0 9 9 1 1 5 5 3 3 9 9 0 0 Rev AB R R e e c c y y c c l l a a b b l l e e p p a a r r t t I I m m p p o o r r t t a a n n t t Q Q u u a a n n t t i i t t y y I I t t e e m m...
Page 41
S S e e c c u u r r e e C C o o n n n n e e c c t t B B B B I I D D N N o o t t e e - - Give this form to your Stryker representative or your IT system analyst so they can create the associations on the server.
Page 43
S S e e c c u u r r e e ® ® C C o o n n n n e e c c t t ™ ™ B B e e d d i i e e n n u u n n g g s s - - u u n n d d W W a a r r t t u u n n g g s s h h a a n n d d b b u u c c h h 521200380100 Connected Hospital®...
Page 45
Seriennummer Europäisches Medizinprodukt CE-Kennzeichnung Bevollmächtigter in der Europäischen Gemeinschaft Für Informationen zu US-Patenten siehe www.stryker.com/patents Hersteller Schutz vor spritzenden Flüssigkeiten I I P P X X 4 4 Elektrische Geräte der Schutzklasse II: Geräte, bei denen der Schutz gegen elektrische Schläge nicht nur von der Basisisolierung abhängt, sondern zusätzliche Sicherheitsvorkehrungen wie...
Page 47
I I n n h h a a l l t t s s v v e e r r z z e e i i c c h h n n i i s s Symbole ................................41 Definition von Warnung/Vorsicht/Hinweis......................2 Zusammenfassung der Sicherheitsvorkehrungen ..................2 Einführung ..............................4 Produktbeschreibung ..........................4...
Page 48
D D e e f f i i n n i i t t i i o o n n v v o o n n W W a a r r n n u u n n g g / / V V o o r r s s i i c c h h t t / / H H i i n n w w e e i i s s Die Begriffe W W A A R R N N U U N N G G , V V O O R R S S I I C C H H T T und H H I I N N W W E E I I S S haben eine besondere Bedeutung und müssen sorgfältig durchgelesen werden.
Page 49
• Alle Produkte nach der Desinfektion stets mit sauberem Wasser (bzw. 70%igem Isopropylalkohol, falls V V i i r r e e x x ® TB verwendet wird) abwischen und trocknen lassen. Manche Desinfektionsmittel sind ätzend und können das Produkt beschädigen.
Page 50
Das Stryker Modell 521200380100 S S e e c c u u r r e e ® C C o o n n n n e e c c t t ™ ist eine kabellose Schwesternruflösung.
Page 51
(1,7 m) von S S e e c c u u r r e e C C o o n n n n e e c c t t entfernt ist. Stryker behält sich das Recht vor, technische Daten ohne Vorankündigung zu ändern.
Page 52
Die Betriebsanleitung bzw. das Wartungshandbuch steht unter der folgenden Adresse im Internet zur Verfügung: https:// techweb.stryker.com/. Bei Anrufen beim Stryker-Kundendienst oder technischen Support bitte die Seriennummer (A) des jeweiligen Stryker- Produkts bereithalten. Seriennummer bei allen schriftlichen Mitteilungen angeben. P P o o s s i i t t i i o o n n d d e e r r S S e e r r i i e e n n n n u u m m m m e e r r Seriennummer (A) und Bettplatz-Identifikationsnummer (Bed Bay Identification Number, BBID) sind auf der Unterseite des Produkts zu finden (Abbildung 1).
Page 53
H H i i n n w w e e i i s s - - Um die Konfiguration der DIP-Schalter zu bestätigen, den Kundendienst bzw. technischen Support von Stryker kontaktieren ( Kontaktinformationen (Seite 6)). A A b b b b i i l l d d u u n n g g 2 2 – – S S c c h h a a l l t t e e r r k k o o n n f f i i g g u u r r a a t t i i o o n n 3.
Page 54
4 in. - 5 in. (10.2 cm - 12.7 cm) ± 0.5 in. (± 1.3 cm) A A b b b b i i l l d d u u n n g g 3 3 – – M M o o n n t t a a g g e e s s p p e e z z i i f f i i k k a a t t i i o o n n e e n n 6.
Page 55
• Wenn ein Produkt zu einem anderen konfigurierten S S e e c c u u r r e e C C o o n n n n e e c c t t verlagert wird, sind keine Änderungen erforderlich, da sich das Produkt automatisch verbindet. 5212-009-205 Rev AA.1...
Page 56
I I n n s s t t a a l l l l a a t t i i o o n n s s - - C C h h e e c c k k l l i i s s t t e e Diese Checkliste für 521200380100 S S e e c c u u r r e e C C o o n n n n e e c c t t befolgen: Bestätigen, dass nach der Installation keine unbenutzten Komponenten übrig bleiben.
Page 57
B B e e t t r r i i e e b b A A n n z z e e i i g g e e n n / / F F u u n n k k t t i i o o n n e e n n v v o o n n S S e e c c u u r r e e C C o o n n n n e e c c t t S S e e c c u u r r e e C C o o n n n n e e c c t t weist Anzeigesymbole und eine Taste zum Zurücksetzen des Alarms für diskonnektiertes Schwesternrufkabel (A) oben auf dem Produkt auf.
Page 58
H H i i n n w w e e i i s s - - Das Stryker Service-Tool ist erforderlich, wenn der S S e e c c u u r r e e C C o o n n n n e e c c t t Scanner verwendet wird.
Page 59
H H i i n n w w e e i i s s - - Das Stryker Service-Tool ist erforderlich, wenn der S S e e c c u u r r e e C C o o n n n n e e c c t t Scanner verwendet wird.
Page 60
Bevor die vorbeugende Wartung vorgenommen wird, muss das Produkt außer Betrieb genommen werden. Alle aufgeführten Punkte müssen bei allen Produkten von Stryker Medical im Rahmen der jährlichen vorbeugenden Wartung überprüft werden. Je nach Ausmaß der Produktnutzung können auch häufigere vorbeugende Wartungsprüfungen erforderlich sein.
Page 61
R R e e i i n n i i g g u u n n g g V V O O R R S S I I C C H H T T • Keine Reinigungs-, Desinfektions-, Instandhaltungs- oder Wartungsarbeiten durchführen, während das Produkt verwendet wird.
Page 62
D D e e s s i i n n f f e e k k t t i i o o n n V V O O R R S S I I C C H H T T •...
Page 63
S S t t ö ö r r u u n n g g s s b b e e h h e e b b u u n n g g L L ö ö s s u u n n g g P P r r o o b b l l e e m m 1.
Page 64
S S e e r r v v i i c c e e A A u u s s t t a a u u s s c c h h d d e e s s A A k k k k u u s s ( ( 3 3 7 7 - - p p o o l l i i g g e e r r V V e e r r b b i i n n d d e e r r , , l l i i n n k k e e S S e e i i t t e e ) ) E E r r f f o o r r d d e e r r l l i i c c h h e e W W e e r r k k z z e e u u g g e e : : •...
Page 65
A A b b b b i i l l d d u u n n g g 8 8 – – A A u u s s t t a a u u s s c c h h d d e e s s A A k k k k u u s s ( ( l l i i n n k k s s ) ) 8.
Page 66
A A u u s s t t a a u u s s c c h h d d e e s s A A k k k k u u s s ( ( 3 3 7 7 - - p p o o l l i i g g e e r r V V e e r r b b i i n n d d e e r r , , r r e e c c h h t t e e S S e e i i t t e e ) ) E E r r f f o o r r d d e e r r l l i i c c h h e e W W e e r r k k z z e e u u g g e e : : •...
Page 67
A A u u f f b b a a u u d d e e s s S S e e c c u u r r e e C C o o n n n n e e c c t t P P o o s s i i t t i i o o n n s s a a n n z z e e i i g g e e r r s s , , l l i i n n k k s s 521200380105 Rev.
Page 68
Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 69
Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 70
A A r r t t i i k k e e l l N N u u m m m m e e r r N N a a m m e e A A n n z z a a h h l l 300900380940 Wandseitige Infrarotplatine 521200380111...
Page 71
A A r r t t i i k k e e l l N N u u m m m m e e r r N N a a m m e e A A n n z z a a h h l l 521200380803 37-poliges Kabel, komplett 521200380806...
Page 72
A A u u f f b b a a u u d d e e s s S S e e c c u u r r e e C C o o n n n n e e c c t t P P o o s s i i t t i i o o n n s s a a n n z z e e i i g g e e r r s s , , r r e e c c h h t t s s 521200380110 Rev.
Page 73
Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 74
Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 75
A A r r t t i i k k e e l l N N u u m m m m e e r r N N a a m m e e A A n n z z a a h h l l 300900380940 Wandseitige Infrarotplatine 521200380111...
Page 76
A A r r t t i i k k e e l l N N u u m m m m e e r r N N a a m m e e A A n n z z a a h h l l 521200380801 Akkukabel, komplett 521200380802...
Page 77
H H i i n n w w e e i i s s e e f f ü ü r r d d r r a a h h t t l l o o s s e e P P r r o o d d u u k k t t e e Für Produkte, die mit der drahtlosen Kommunikationstechnologie ausgestattet sind, gelten die folgenden Hinweise für das jeweils angegebene Land: L L a a n n d d...
Page 78
E E M M V V - - I I n n f f o o r r m m a a t t i i o o n n e e n n W W A A R R N N U U N N G G •...
Page 79
L L e e i i t t l l i i n n i i e e u u n n d d H H e e r r s s t t e e l l l l e e r r e e r r k k l l ä ä r r u u n n g g – – e e l l e e k k t t r r o o m m a a g g n n e e t t i i s s c c h h e e S S t t ö ö r r f f e e s s t t i i g g k k e e i i t t Der Fußboden sollte aus Holz, Beton oder Keramikfliesen bestehen.
Page 80
Tragbare und mobile HF- Kommunikationsgeräte sollten entsprechend den Angaben in der Tabelle „Empfohlene Schutzabstände zwischen tragbaren und mobilen HF- Kommunikationsgeräten und dem 521200380100 S S e e c c u u r r e e C C o o n n n n e e c c t t Positionsanzeiger“...
Page 81
Feldstärken von stationären Sendern wie Basisstationen von Funktelefonen (Mobiltelefon/schnurloses Telefon) und mobilen Landfunkdiensten, Amateurfunkgeräten, AM- und FM-Radiosendern und TV-Sendern, lassen sich theoretisch nicht genau vorherbestimmen. Zur Beurteilung der durch stationäre HF-Sender geschaffenen elektromagnetischen Umgebung sollte eine elektromagnetische Standortvermessung erwogen werden. Wenn die gemessene Feldstärke am Anwendungsort des 521200380100 S S e e c c u u r r e e C C o o n n n n e e c c t t Positionsanzeigers den oben erwähnten zutreffenden HF- Konformitätspegel übersteigt, sollteder 521200380100 S S e e c c u u r r e e C C o o n n n n e e c c t t Positionsanzeiger beobachtet werden, um die normale Funktion zu bestätigen.
Page 82
R R e e c c y y c c l l i i n n g g p p a a s s s s 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 0 0 5 5 / / 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 1 1 0 0 Rev.
Page 83
7 7 0 0 0 0 0 0 0 0 0 0 8 8 8 8 6 6 7 7 9 9 5 5 Rev. AB N N u u m m m m e e r r d d e e s s r r e e c c y y c c l l i i n n g g f f ä...
Page 84
7 7 0 0 0 0 0 0 0 0 0 0 9 9 1 1 5 5 3 3 9 9 0 0 Rev. AB N N u u m m m m e e r r d d e e s s r r e e c c y y c c l l i i n n g g f f ä...
Page 85
Z Z i i m m m m e e r r n n u u m m m m e e r r / / S S t t a a n n d d o o r r t t H H i i n n w w e e i i s s - - Dieses Formular dem zuständigen Stryker-Vertreter oder IT-Systemanalysten aushändigen, damit diese Person die Zuweisungen auf dem Server vornehmen kann.
Page 87
S S e e c c u u r r e e ® ® C C o o n n n n e e c c t t ™ ™ M M a a n n u u a a l l d d e e u u s s o o y y m m a a n n t t e e n n i i m m i i e e n n t t o o 521200380100 Connected Hospital®...
Page 89
Número de serie Producto sanitario europeo Marca CE Representante autorizado en la Comunidad Europea Para ver las patentes estadounidenses, visite www.stryker.com/patents Fabricante Protección contra salpicaduras líquidas I I P P X X 4 4 Equipo eléctrico de clase II: equipo en el que la protección contra las descargas eléctricas no consiste solamente en un aislamiento básico, sino que incluye precauciones de seguridad...
Page 91
Í Í n n d d i i c c e e Símbolos ................................43 Definición de advertencia, precaución y nota ....................2 Resumen de las precauciones de seguridad ....................2 Introducción ..............................3 Descripción del producto ........................3 Vida útil prevista.............................3 Eliminación/reciclado ........................3 Contraindicaciones..........................3 Especificaciones ............................3 Especificaciones de la radio por Bluetooth..................4 Información de contacto..........................5...
Page 92
D D e e f f i i n n i i c c i i ó ó n n d d e e a a d d v v e e r r t t e e n n c c i i a a , , p p r r e e c c a a u u c c i i ó ó n n y y n n o o t t a a Las palabras A A D D V V E E R R T T E E N N C C I I A A , P P R R E E C C A A U U C C I I Ó...
Page 93
El S S e e c c u u r r e e ® C C o o n n n n e e c c t t ™ modelo 521200380100 de Stryker es una solución inalámbrica...
Page 94
5,5 pies (1,7 m) del S S e e c c u u r r e e C C o o n n n n e e c c t t . Stryker se reserva el derecho a cambiar las especificaciones sin previo aviso.
Page 95
Para ver en línea el manual de uso o de mantenimiento de su producto, visite https://techweb.stryker.com/. Tenga a mano el número de serie (A) del producto de Stryker cuando llame al Servicio de Atención al Cliente o al Servicio de Asistencia Técnica de Stryker. Incluya el número de serie en todas las comunicaciones escritas.
Page 96
N N o o t t a a - - Para confirmar la configuración del interruptor DIP, póngase en contacto con el servicio de atención al cliente o el servicio técnico de Stryker ( Información de contacto (página 5)). F F i i g g u u r r a a 2 2 – – C C o o n n f f i i g g u u r r a a c c i i ó ó n n d d e e l l i i n n t t e e r r r r u u p p t t o o r r 3.
Page 97
4 in. - 5 in. (10.2 cm - 12.7 cm) ± 0.5 in. (± 1.3 cm) F F i i g g u u r r a a 3 3 – – E E s s p p e e c c i i f f i i c c a a c c i i o o n n e e s s d d e e m m o o n n t t a a j j e e 6.
Page 98
• Si el producto se lleva a otro S S e e c c u u r r e e C C o o n n n n e e c c t t que ya esté configurado, no es necesario hacer cambios, ya que el producto se conectará...
Page 99
L L i i s s t t a a d d e e c c o o m m p p r r o o b b a a c c i i ó ó n n d d e e l l a a i i n n s s t t a a l l a a c c i i ó ó n n Siga esta lista de verificación para el S S e e c c u u r r e e C C o o n n n n e e c c t t 521200380100: Confirme que no le quede ningún componente sin usar después de la instalación.
Page 100
F F u u n n c c i i o o n n a a m m i i e e n n t t o o I I n n d d i i c c a a d d o o r r e e s s y y f f u u n n c c i i o o n n e e s s d d e e S S e e c c u u r r e e C C o o n n n n e e c c t t El S S e e c c u u r r e e C C o o n n n n e e c c t t tiene símbolos indicadores y un botón para cancelar la llamada al personal de enfermería (A), ubicado en la parte superior del producto.
Page 101
N N o o t t a a - - Se necesita la herramienta de servicio técnico de Stryker si utiliza el escáner S S e e c c u u r r e e C C o o n n n n e e c c t t .
Page 102
N N o o t t a a - - Se necesita la herramienta de servicio técnico de Stryker si utiliza el escáner S S e e c c u u r r e e C C o o n n n n e e c c t t .
Page 103
Retire el producto del servicio antes de realizar la inspección de mantenimiento preventivo. Revise todos los componentes incluidos en la lista durante el mantenimiento preventivo anual de todos los productos de Stryker Medical. Es posible que deba realizar revisiones de mantenimiento preventivo con más frecuencia en función de su nivel de uso del producto. Las reparaciones solo puede realizarlas personal cualificado.
Page 104
L L i i m m p p i i e e z z a a P P R R E E C C A A U U C C I I Ó Ó N N • No limpie, desinfecte, repare ni realice tareas de mantenimiento en el producto mientras se esté utilizando. •...
Page 105
D D e e s s i i n n f f e e c c c c i i ó ó n n P P R R E E C C A A U U C C I I Ó Ó N N •...
Page 106
R R e e s s o o l l u u c c i i ó ó n n d d e e p p r r o o b b l l e e m m a a s s P P r r o o b b l l e e m m a a S S o o l l u u c c i i ó...
Page 107
M M a a n n t t e e n n i i m m i i e e n n t t o o S S u u s s t t i i t t u u c c i i ó ó n n d d e e l l a a b b a a t t e e r r í í a a ( ( c c o o n n e e c c t t o o r r d d e e 3 3 7 7 p p i i n n e e s s , , l l a a z z o o i i z z q q u u i i e e r r d d o o ) ) H H e e r r r r a a m m i i e e n n t t a a s s n n e e c c e e s s a a r r i i a a s s : : •...
Page 108
F F i i g g u u r r a a 8 8 – – S S u u s s t t i i t t u u c c i i ó ó n n d d e e l l a a b b a a t t e e r r í í a a ( ( i i z z q q u u i i e e r r d d a a ) ) 8.
Page 109
S S u u s s t t i i t t u u c c i i ó ó n n d d e e l l a a b b a a t t e e r r í í a a ( ( c c o o n n e e c c t t o o r r d d e e 3 3 7 7 p p i i n n e e s s , , l l a a z z o o d d e e r r e e c c h h o o ) ) H H e e r r r r a a m m i i e e n n t t a a s s n n e e c c e e s s a a r r i i a a s s : : •...
Page 110
M M o o n n t t a a j j e e d d e e l l l l o o c c a a l l i i z z a a d d o o r r S S e e c c u u r r e e C C o o n n n n e e c c t t , , i i z z q q u u i i e e r r d d a a 521200380105 Rev.
Page 111
Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 112
Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 113
E E l l e e m m e e n n t t o o N N ú ú m m e e r r o o N N o o m m b b r r e e C C a a n n t t i i d d a a d d 300900380940 Placa de infrarrojos de la pared...
Page 114
E E l l e e m m e e n n t t o o N N ú ú m m e e r r o o N N o o m m b b r r e e C C a a n n t t i i d d a a d d 521200380802 Conjunto de cables de la placa de...
Page 115
M M o o n n t t a a j j e e d d e e l l l l o o c c a a l l i i z z a a d d o o r r S S e e c c u u r r e e C C o o n n n n e e c c t t , , d d e e r r e e c c h h a a 521200380110 Rev.
Page 116
Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 117
Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 118
E E l l e e m m e e n n t t o o N N ú ú m m e e r r o o N N o o m m b b r r e e C C a a n n t t i i d d a a d d 300900380940 Placa de infrarrojos de la pared...
Page 119
E E l l e e m m e e n n t t o o N N ú ú m m e e r r o o N N o o m m b b r r e e C C a a n n t t i i d d a a d d 521200380121 Placa de montaje...
Page 120
N N o o t t i i f f i i c c a a c c i i o o n n e e s s i i n n a a l l á á m m b b r r i i c c a a s s Para los productos que cuentan con tecnología de comunicación inalámbrica, estas declaraciones se aplican a los países de la siguiente manera: P P a a í...
Page 121
I I n n f f o o r r m m a a c c i i ó ó n n d d e e c c o o m m p p a a t t i i b b i i l l i i d d a a d d e e l l e e c c t t r r o o m m a a g g n n é é t t i i c c a a A A D D V V E E R R T T E E N N C C I I A A •...
Page 122
G G u u í í a a y y d d e e c c l l a a r r a a c c i i ó ó n n d d e e l l f f a a b b r r i i c c a a n n t t e e s s o o b b r r e e i i n n m m u u n n i i d d a a d d e e l l e e c c t t r r o o m m a a g g n n é é t t i i c c a a La calidad de la red ±2 kV para las líneas de ±2 kV para las líneas de...
Page 123
Los equipos portátiles y móviles de comunicación por radiofrecuencia deberán seguir las orientaciones indicadas en la tabla «Distancias de separación recomendadas entre los equipos portátiles y móviles de comunicación por radiofrecuencia y el localizador S S e e c c u u r r e e C C o o n n n n e e c c t t 521200380100».
Las intensidades de campo de transmisores fijos, como las estaciones base de radioteléfonos (móviles e inalámbricos) y radiotransmisores terrestres, los equipos de radioaficionados, las emisoras de radio AM y FM, y las emisoras de televisión, no pueden predecirse de forma teórica con exactitud. Para evaluar el entorno electromagnético generado por transmisores de radiofrecuencia fijos, debe considerarse la posibilidad de hacer un estudio electromagnético del lugar.
Page 125
P P a a s s a a p p o o r r t t e e d d e e r r e e c c i i c c l l a a j j e e 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 0 0 5 5 / / 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 1 1 0 0 Rev.
Page 126
7 7 0 0 0 0 0 0 0 0 0 0 8 8 8 8 6 6 7 7 9 9 5 5 Rev. AB I I n n f f o o r r m m a a c c i i ó ó n n N N ú...
Page 127
7 7 0 0 0 0 0 0 0 0 0 0 9 9 1 1 5 5 3 3 9 9 0 0 Rev. AB I I n n f f o o r r m m a a c c i i ó ó n n N N ú...
Page 128
N N ú ú m m e e r r o o o o u u b b i i c c a a c c i i ó ó n n d d e e l l a a h h a a b b i i t t a a c c i i ó ó n n N N o o t t a a - - Entregue este formulario a su representante de Stryker o al analista del sistema informático para que puedan crear las asociaciones en el servidor.
Page 129
S S e e c c u u r r e e ® ® C C o o n n n n e e c c t t ™ ™ M M a a n n u u e e l l d d ’ ’ u u t t i i l l i i s s a a t t i i o o n n e e t t d d ’ ’ e e n n t t r r e e t t i i e e n n 521200380100 Connected Hospital®...
Page 131
Numéro de série Dispositif médical européen Marquage CE Mandataire établi dans la Communauté européenne Pour les brevets américains, consulter www.stryker.com/patents Fabricant Protection contre les projections de liquides I I P P X X 4 4 Équipement électrique de classe II : équipement pour lequel la protection contre les chocs électriques ne repose pas uniquement sur une isolation de base, mais qui comprend des...
Page 133
T T a a b b l l e e d d e e s s m m a a t t i i è è r r e e s s Symboles................................41 Définition de « Avertissement », « Mise en garde » et « Remarque » ..............2 Résumé...
Page 134
D D é é f f i i n n i i t t i i o o n n d d e e « « A A v v e e r r t t i i s s s s e e m m e e n n t t » » , , « « M M i i s s e e e e n n g g a a r r d d e e » » e e t t « « R R e e m m a a r r q q u u e e » » Les termes A A V V E E R R T T I I S S S S E E M M E E N N T T , M M I I S S E E E E N N G G A A R R D D E E et R R E E M M A A R R Q Q U U E E ont une signification particulière et doivent faire l’objet d’une lecture attentive.
Page 135
• Toujours nettoyer le produit à l’eau claire (ou avec de l'alcool isopropylique à 70 %, en cas d'utilisation de V V i i r r e e x x ® TB) et sécher après la désinfection. Certains produits désinfectants sont de nature corrosive et peuvent endommager le produit.
Page 136
Le S S e e c c u u r r e e ® C C o o n n n n e e c c t t ™ modèle 521200380100 de Stryker est une solution d’appel infirmier sans câle.
Page 137
C C o o n n n n e e c c t t . Stryker se réserve le droit de modifier ces caractéristiques sans préavis. Les caractéristiques techniques indiquées sont approximatives et susceptibles de varier légèrement d’un système à l’autre ou en fonction des fluctuations de l’alimentation électrique.
Page 138
Pour consulter votre mode d'emploi ou votre manuel d'entretien en ligne, consulter https://techweb.stryker.com/. Avoir le numéro de série (A) du produit Stryker à disposition avant d’appeler le service clientèle ou le support technique de Stryker. Inclure le numéro de série dans toutes les communications écrites.
Page 139
R R e e m m a a r r q q u u e e - - Pour confirmer la configuration des commutateurs DIP, contacter le service clients ou le service technique Stryker ( Coordonnées (page 6)). F F i i g g u u r r e e 2 2 – – C C o o n n f f i i g g u u r r a a t t i i o o n n d d e e s s c c o o m m m m u u t t a a t t e e u u r r s s 3.
Page 140
4 in. - 5 in. (10.2 cm - 12.7 cm) ± 0.5 in. (± 1.3 cm) F F i i g g u u r r e e 3 3 – – C C a a r r a a c c t t é é r r i i s s t t i i q q u u e e s s t t e e c c h h n n i i q q u u e e s s d d e e m m o o n n t t a a g g e e 6.
Page 141
• Si un produit est transféré à un autre S S e e c c u u r r e e C C o o n n n n e e c c t t configuré, aucune modification n'est nécessaire car le produit se connecte automatiquement.
Page 142
L L i i s s t t e e d d e e c c o o n n t t r r ô ô l l e e d d ’ ’ i i n n s s t t a a l l l l a a t t i i o o n n Suivre cette liste de contrôle pour le S S e e c c u u r r e e C C o o n n n n e e c c t t 521200380100 : Confirmer qu’il ne reste aucun composant non utilisé...
F F o o n n c c t t i i o o n n n n e e m m e e n n t t I I n n d d i i c c a a t t e e u u r r s s / / f f o o n n c c t t i i o o n n s s d d u u S S e e c c u u r r e e C C o o n n n n e e c c t t Le S S e e c c u u r r e e C C o o n n n n e e c c t t comporte des pictogrammes lumineux et un bouton d’annulation en cas de débranchement du câble de communication pour appel infirmier (A) qui est situé...
Page 144
R R e e m m a a r r q q u u e e - - Outil d’entretien Stryker requis en cas d’utilisation du scanner S S e e c c u u r r e e C C o o n n n n e e c c t t .
Page 145
R R e e m m a a r r q q u u e e - - Outil d’entretien Stryker requis en cas d’utilisation du scanner S S e e c c u u r r e e C C o o n n n n e e c c t t .
Page 146
Mettre le produit hors service avant d’effectuer les vérifications d’entretien préventif. Vérifier tous les éléments mentionnés pendant l’entretien préventif annuel pour tous les produits Stryker Medical. Il peut être nécessaire d’effectuer les vérifications d’entretien préventif plus fréquemment en fonction du degré d’utilisation du produit. Toute réparation doit être effectuée exclusivement par du personnel qualifié.
Page 147
N N e e t t t t o o y y a a g g e e M M I I S S E E E E N N G G A A R R D D E E •...
Page 148
D D é é s s i i n n f f e e c c t t i i o o n n M M I I S S E E E E N N G G A A R R D D E E •...
D D é é p p a a n n n n a a g g e e P P r r o o b b l l è è m m e e S S o o l l u u t t i i o o n n 1.
Page 150
E E n n t t r r e e t t i i e e n n R R e e m m p p l l a a c c e e m m e e n n t t d d e e l l a a b b a a t t t t e e r r i i e e ( ( c c o o n n n n e e c c t t e e u u r r à à 3 3 7 7 b b r r o o c c h h e e s s , , c c ô ô t t é é g g a a u u c c h h e e ) ) O O u u t t i i l l s s r r e e q q u u i i s s : : •...
Page 151
F F i i g g u u r r e e 8 8 – – R R e e m m p p l l a a c c e e m m e e n n t t d d e e l l a a b b a a t t t t e e r r i i e e ( ( c c ô ô t t é é g g a a u u c c h h e e ) ) 8.
Page 152
R R e e m m p p l l a a c c e e m m e e n n t t d d e e l l a a b b a a t t t t e e r r i i e e ( ( c c o o n n n n e e c c t t e e u u r r à à 3 3 7 7 b b r r o o c c h h e e s s , , c c ô ô t t é é d d r r o o i i t t ) ) O O u u t t i i l l s s r r e e q q u u i i s s : : •...
Page 153
L L o o c c a a l l i i s s a a t t e e u u r r S S e e c c u u r r e e C C o o n n n n e e c c t t , , c c ô ô t t é é g g a a u u c c h h e e 521200380105 Rev AC (pour référence uniquement) Torque item AB to 0.34 ±...
Page 154
Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 155
Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 156
N N u u m m é é r r o o N N o o m m Q Q u u a a n n t t i i t t é é É É l l é é m m e e n n t t 300900380940 Circuit IR côté...
Page 157
N N u u m m é é r r o o N N o o m m Q Q u u a a n n t t i i t t é é É É l l é é m m e e n n t t 521200380803 Câble avec connecteur à...
Page 158
L L o o c c a a l l i i s s a a t t e e u u r r S S e e c c u u r r e e C C o o n n n n e e c c t t , , c c ô ô t t é é d d r r o o i i t t 521200380110 Rev AC (pour référence uniquement) Torque item AB Torque item AD...
Page 159
Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 160
Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 161
N N u u m m é é r r o o N N o o m m Q Q u u a a n n t t i i t t é é É É l l é é m m e e n n t t 300900380940 Circuit IR côté...
Page 162
N N u u m m é é r r o o N N o o m m Q Q u u a a n n t t i i t t é é É É l l é é m m e e n n t t 521200380801 Câble de la batterie 521200380802...
Page 163
N N o o t t i i f f i i c c a a t t i i o o n n s s s s a a n n s s f f i i l l Pour les produits équipés d’une technologie de communication sans fil, ces notifications s’appliquent aux pays indiqués ci- dessous : P P a a y y s s...
Page 164
I I n n f f o o r r m m a a t t i i o o n n s s d d e e C C E E M M A A V V E E R R T T I I S S S S E E M M E E N N T T •...
Page 165
D D i i r r e e c c t t i i v v e e s s e e t t d d é é c c l l a a r r a a t t i i o o n n d d u u f f a a b b r r i i c c a a n n t t – – I I m m m m u u n n i i t t é é é é l l e e c c t t r r o o m m a a g g n n é é t t i i q q u u e e La qualité...
Page 166
Les équipements de communication RF portables et mobiles doivent respecter les indications du tableau intitulé « Distances de séparation recommandées entre les équipements de communication RF portables et mobiles et le localisateur S S e e c c u u r r e e C C o o n n n n e e c c t t 521200380100 ».
Les intensités de champ des émetteurs fixes, tels que les stations de base pour les radiotéléphones (cellulaires/sans fil) et les radios mobiles terrestres, la radio amateur, la radiodiffusion AM et FM, et la diffusion télévisuelle, ne peuvent pas être prévues théoriquement avec exactitude. Pour évaluer l’environnement électromagnétique dû aux émetteurs RF fixes, il convient d’envisager un relevé...
Page 168
P P a a s s s s e e p p o o r r t t d d e e r r e e c c y y c c l l a a g g e e 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 0 0 5 5 / / 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 1 1 0 0 Rév.
7 7 0 0 0 0 0 0 0 0 0 0 8 8 8 8 6 6 7 7 9 9 5 5 Rév. AB N N u u m m é é r r o o d d e e p p i i è è c c e e I I n n f f o o r r m m a a t t i i o o n n s s r r e e c c y y c c l l a a b b l l e e i i m m p p o o r r t t a a n n t t e e s s...
Page 170
7 7 0 0 0 0 0 0 0 0 0 0 9 9 1 1 5 5 3 3 9 9 0 0 Rév. AB N N u u m m é é r r o o d d e e p p i i è è c c e e I I n n f f o o r r m m a a t t i i o o n n s s r r e e c c y y c c l l a a b b l l e e i i m m p p o o r r t t a a n n t t e e s s...
Page 171
S S e e c c u u r r e e C C o o n n n n e e c c t t R R e e m m a a r r q q u u e e - - Remettre ce formulaire à un représentant Stryker ou à un analyste informatique afin qu'il puisse créer les associations sur le serveur.
Page 173
S S e e c c u u r r e e ® ® C C o o n n n n e e c c t t ™ ™ M M a a n n u u a a l l d d e e f f u u n n c c i i o o n n a a m m e e n n t t o o / / m m a a n n u u t t e e n n ç ç ã ã o o 521200380100 Connected Hospital®...
Page 175
Número de série Dispositivo médico Europeu Marcação CE Mandatário autorizado na Comunidade Europeia Para patentes dos EUA, consulte www.stryker.com/patents Fabricante Proteção contra salpicos de líquidos I I P P X X 4 4 Equipamento elétrico de Classe II: equipamento cuja proteção contra choques elétricos não depende unicamente de isolamento básico, mas no qual foram previstas precauções adicionais...
Page 177
Í Í n n d d i i c c e e Símbolos ................................43 Definição de Advertência/Precaução/Nota ......................2 Resumo das precauções de segurança ....................2 Introdução ..............................3 Descrição do produto ..........................3 Vida útil prevista.............................3 Eliminação/reciclagem ........................3 Contraindicações ...........................3 Especificações............................3 Especificações de rádio Bluetooth ....................4 Informações para contacto........................5 Localização do número de série .......................5 Instalação ..............................6...
Page 178
D D e e f f i i n n i i ç ç ã ã o o d d e e A A d d v v e e r r t t ê ê n n c c i i a a / / P P r r e e c c a a u u ç ç ã ã o o / / N N o o t t a a Os termos A A D D V V E E R R T T Ê...
Page 179
O S S e e c c u u r r e e ® C C o o n n n n e e c c t t ™ Stryker Modelo 521200380100 é uma solução de chamada de enfermeiro sem fios.
Page 180
S S e e c c u u r r e e C C o o n n n n e e c c t t . A Stryker reserva-se o direito de alterar as especificações sem aviso prévio.
Page 181
Para consultar o manual de operações ou manutenção online, vá a https://techweb.stryker.com/. Tenha o número de série (A) do seu produto da Stryker à mão quando telefonar para o Serviço de Apoio ao Cliente ou Assistência Técnica da Stryker. Inclua o número de série em toda a comunicação escrita.
Page 182
N N o o t t a a - - Para confirmar a configuração dos comutadores dip, contacte o serviço de apoio a clientes Stryker ou o apoio técnico ( Informações para contacto (página 5)).
Page 183
4 in. - 5 in. (10.2 cm - 12.7 cm) ± 0.5 in. (± 1.3 cm) F F i i g g u u r r a a 3 3 – – E E s s p p e e c c i i f f i i c c a a ç ç õ õ e e s s d d e e m m o o n n t t a a g g e e m m 6.
Page 184
L L i i s s t t a a d d e e v v e e r r i i f f i i c c a a ç ç ã ã o o d d a a i i n n s s t t a a l l a a ç ç ã ã o o Siga esta lista de verificação para o S S e e c c u u r r e e C C o o n n n n e e c c t t 521200380100: Confirme que não tem quaisquer componentes não utilizados após a instalação.
Page 185
F F u u n n c c i i o o n n a a m m e e n n t t o o I I n n d d i i c c a a d d o o r r e e s s / / f f u u n n ç ç õ õ e e s s d d o o S S e e c c u u r r e e C C o o n n n n e e c c t t O S S e e c c u u r r e e C C o o n n n n e e c c t t tem símbolos indicadores e um cabo de chamada de enfermeiro com botão de cancelamento (A) localizado no topo do produto.
Page 186
N N o o t t a a - - A ferramenta de assistência Stryker é necessária se utilizar o digitalizador S S e e c c u u r r e e C C o o n n n n e e c c t t .
Page 187
N N o o t t a a - - A ferramenta de assistência Stryker é necessária se utilizar o digitalizador S S e e c c u u r r e e C C o o n n n n e e c c t t .
Page 188
Retire o produto de utilização antes da realização da inspeção de manutenção preventiva. Verifique todos os itens indicados durante a manutenção preventiva anual para todos os produtos da Stryker Medical. Poderá ter de realizar verificações da manutenção preventiva mais frequentemente, com base no nível de utilização do produto. A assistência deve ser feita apenas por pessoal qualificado.
Page 189
L L i i m m p p e e z z a a P P R R E E C C A A U U Ç Ç Ã Ã O O • Não proceda à limpeza, à desinfeção, à reparação ou à manutenção enquanto o produto estiver a ser utilizado. •...
Page 190
D D e e s s i i n n f f e e ç ç ã ã o o P P R R E E C C A A U U Ç Ç Ã Ã O O • Não proceda à limpeza, à desinfeção, à reparação ou à manutenção enquanto o produto estiver a ser utilizado. •...
Page 191
R R e e s s o o l l u u ç ç ã ã o o d d e e p p r r o o b b l l e e m m a a s s S S o o l l u u ç...
Page 192
A A s s s s i i s s t t ê ê n n c c i i a a S S u u b b s s t t i i t t u u i i ç ç ã ã o o d d a a b b a a t t e e r r i i a a ( ( c c o o n n e e c c t t o o r r d d e e 3 3 7 7 p p i i n n o o s s , , l l a a d d o o e e s s q q u u e e r r d d o o ) ) F F e e r r r r a a m m e e n n t t a a s s n n e e c c e e s s s s á...
Page 193
F F i i g g u u r r a a 8 8 – – S S u u b b s s t t i i t t u u i i r r a a b b a a t t e e r r i i a a ( ( e e s s q q u u e e r r d d a a ) ) 8.
Page 194
S S u u b b s s t t i i t t u u i i ç ç ã ã o o d d a a b b a a t t e e r r i i a a ( ( c c o o n n e e c c t t o o r r d d e e 3 3 7 7 p p i i n n o o s s , , l l a a d d o o d d i i r r e e i i t t o o ) ) F F e e r r r r a a m m e e n n t t a a s s n n e e c c e e s s s s á...
Page 195
C C o o n n j j u u n n t t o o d d o o l l o o c c a a l l i i z z a a d d o o r r S S e e c c u u r r e e C C o o n n n n e e c c t t , , e e s s q q u u e e r r d d a a 521200380105 Rev AC (apenas referência) Torque item AB to 0.34 ±...
Page 196
Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 197
Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 198
I I t t e e m m N N ú ú m m e e r r o o N N o o m m e e Q Q u u a a n n t t i i d d a a d d e e 300900380940 Placa de infravermelhos da parede 521200380111...
Page 199
I I t t e e m m N N ú ú m m e e r r o o N N o o m m e e Q Q u u a a n n t t i i d d a a d d e e 521200380803 Conjunto do cabo de 37 pinos 521200380806...
Page 200
C C o o n n j j u u n n t t o o d d o o l l o o c c a a l l i i z z a a d d o o r r S S e e c c u u r r e e C C o o n n n n e e c c t t , , d d i i r r e e i i t t a a 521200380110 Rev AC (apenas referência) Torque item AB Torque item AD...
Page 201
Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 202
Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-205 Rev AA.1...
Page 203
I I t t e e m m N N ú ú m m e e r r o o N N o o m m e e Q Q u u a a n n t t i i d d a a d d e e 300900380940 Placa de infravermelhos da parede 521200380111...
Page 204
I I t t e e m m N N ú ú m m e e r r o o N N o o m m e e Q Q u u a a n n t t i i d d a a d d e e 521200380801 Conjunto do cabo da bateria 521200380802...
Page 205
N N o o t t i i f f i i c c a a ç ç õ õ e e s s s s e e m m f f i i o o s s Para produtos equipados com tecnologia de comunicação sem fios, estas afirmações aplicam-se aos países conforme indicado: N N o o t t i i f f i i c c a a ç...
Page 206
I I n n f f o o r r m m a a ç ç õ õ e e s s d d e e C C E E M M A A D D V V E E R R T T Ê Ê N N C C I I A A •...
Page 207
O O r r i i e e n n t t a a ç ç õ õ e e s s e e d d e e c c l l a a r r a a ç ç ã ã o o d d o o f f a a b b r r i i c c a a n n t t e e – – i i m m u u n n i i d d a a d d e e e e l l e e t t r r o o m m a a g g n n é é t t i i c c a a A qualidade da rede elétrica ±2 kV para linhas de ±2 kV para linhas de...
Page 208
Os equipamentos portáteis e móveis de comunicações RF devem seguir as orientações na tabela intitulada “Distâncias de separação recomendadas entre os equipamentos de comunicação RF portáteis e móveis e o localizador S S e e c c u u r r e e C C o o n n n n e e c c t t 521200380100.”...
Page 209
D D i i s s t t â â n n c c i i a a s s d d e e s s e e p p a a r r a a ç ç ã ã o o r r e e c c o o m m e e n n d d a a d d a a s s e e n n t t r r e e e e q q u u i i p p a a m m e e n n t t o o d d e e c c o o m m u u n n i i c c a a ç ç ã ã o o p p o o r r R R F F , , p p o o r r t t á á t t i i l l e e m m ó ó v v e e l l , , e e o o l l o o c c a a l l i i z z a a d d o o r r S S e e c c u u r r e e C C o o n n n n e e c c t t 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 0 0 0 0 O localizador S S e e c c u u r r e e C C o o n n n n e e c c t t 521200380100 destina-se a ser utilizada num ambiente eletromagnético em que as perturbações por RF radiada estejam controladas.
Page 210
P P a a s s s s a a p p o o r r t t e e d d e e r r e e c c i i c c l l a a g g e e m m 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 0 0 5 5 / / 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 1 1 0 0 Rev AC I I n n f f o o r r m m a a ç...
Page 211
7 7 0 0 0 0 0 0 0 0 0 0 8 8 8 8 6 6 7 7 9 9 5 5 Rev AB I I n n f f o o r r m m a a ç ç õ õ e e s s N N ú...
Page 212
7 7 0 0 0 0 0 0 0 0 0 0 9 9 1 1 5 5 3 3 9 9 0 0 Rev AB I I n n f f o o r r m m a a ç ç õ õ e e s s N N ú...
Page 213
B B B B I I D D d d o o S S e e c c u u r r e e C C o o n n n n e e c c t t N N o o t t a a - - Entregue este formulário ao seu mandatário Stryker ou ao seu analista de sistemas de TI de forma a que possam criar as associações no servidor.
Page 214
Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 5212-009-205 Rev AA.1 2021/01 WCR: AA.1...