Maintenance
6.0 Maintenance
When using the instrument in compliance with the in-
struction manual no special maintenance is required.
6.1 Cleaning
If the instrument is dirty after usage, it is advised to clean
it by using a humid cloth and mild household detergents.
Never use acid detergents or dissolvants.
6.2 Fuse replacement (UNIdreh EVU only)
The UNIdreh EVU has fixed measuring lines and test
probes with built-in fuses 0,5 A / 500 V, 6 x 32 mm.
The front part of the test probe is screwed together with
the handheld part. The fuse can be taken after solving the
bolted joint.
Ensure that the fuses are in perfect condition and cor-
rectly positioned in fuse holder.
Fuse: 500 mA / 500 V, 56 kA cut-off capacity, 6,3 x 32 mm
In order to avoid fire hazard only use the correct fuse
with characteristics regarding type, voltage and cut
off as specified in the instruction manual. The use of
auxiliary fuses especially short-circuiting of fuse
holders is strictly forbidden.