Technical information
Battery:
RENATA type CR2450N mercury-free button-cell battery for
approx. 500 measurements, battery to be replaced only by UEBE
Medical GmbH. The unit may only be repaired by the manufac-
turer or by an expressly authorised organisation. Do not open
the unit, as this will invalidate the warranty.
Service:
Should your unit show evidence of a fault, we ask you not to
attempt repairs yourself, but instead to send the unit with a
short description of the fault directly to UEBE Medical GmbH for
inspection, as unit parts cannot be repaired individually.
Calibration:
The unit does not have to be calibrated; the metrological accu-
racy of the unit will be automatically checked when the battery
is replaced by UEBE Medical GmbH.
Protection class IP:
IP 20: Protected against solid foreign particles with a diameter
of more than 12.5 mm, no protection against water.
Expected service
5 years
life:
Additional equipment connected to medical electrical equipment
must comply with the respective IEC or ISO standards (e.g. IEC 60950
for data-processing equipment). Furthermore, all configurations shall
comply with the normative requirements for medical systems (see IEC
60601-1-1 or Section 16 of the 3rd edition of IEC 60601-1, respec-
tively). Anybody connecting additional equipment to medical electri-
cal equipment is a system configurer and is therefore responsible for
ensuring that the system complies with the normative requirements
for systems. Attention is drawn to the fact that local laws take priority
over the above-mentioned normative requirements. If in doubt, con-
sult your local specialised dealer or the Technical Service department.
EN-80

General provisions

Explanation of symbols

Degree of protection against electric shock: TYP BF
Please observe operating instructions!
0123
This product complies with the Council Directive 93/42/
EC from 5 September 2007 regarding medical devices
and bears the mark CE 0123 (TÜV SÜD Product Service
GmbH).
50 °C
Storage and transportation conditions
Ambient temperature -10 to +50 °C
-10 °C
93%
Protect against moisture/humidity
Relative air humidity up to 93 %
Keep dry
Manufacturer

Disposal

Batteries and technical appliances must not be dis-
posed of with domestic waste, but should be handed in
at the appropriate collection and disposal points.
.........................
.........................