Conditioning, calibration, measurement
General
information
Before measuring
18
4
When doing so refill any spilled bridge electrolyte so that the inner
junctions are covered with bridge electrolyte.
5
Rinse the combination electrode with deionized water.
6
Wipe the shaft using a clean paper towel.
Conditioning, calibration, measurement
We recommend using the cupric electrodes only for the end point
determination of titration procedures.
When operating the electrode ensure that
 the filling opening (1) for the bridge electrolyte is open
 the inner junctions (3) are covered with bridge electrolyte
 no air bubbles are in the bridge electrolyte
 the depth of immersion is within the optimum range:
Minimum depth of im-
mersion
Maximum depth of im-
mersion
1
Condition every ion sensitive electrode in standard solution before
using it.
1
2
3
4
The ground junction (4) must be covered
Approx. 1 cm below the fluid level (2) of the
bridge electrolyte
Cu 800
ba15357d04
12/2021