Télécharger Imprimer la page

WIKA 532.54.100 Mode D'emploi page 14

Publicité

Les langues disponibles

Les langues disponibles

2. Safety
10. Assessment of all accessory components
All accessories (e.g. valves or attachment components) must be assessed in
combination with the delivered instruments by the end user. Particularly the requi-
rements of grounding and prevention of electrostatic charges must be considered.
EN
11. Suitable protective measures
The operator must recognise ignition hazards and take suitable protective measu-
res. See chapter 2.2 "Responsibility of the operator".
12. Description of the protection level
The operator must understand the Ex marking based on the description in chapter
2.6 "Labelling / safety marks" and observe it for their application.
13. Checking of the instrument labelling
The legibility of the marking must be observed during time in use but at least
during inspection periods of three years. See chapter 2.2 "Responsibility of the
operator".
14. Instruments with mark pointer
For instruments with marking pointer, ensure that there are no electrostatic
charging mechanisms at the marking pointer.
15. Protection against external impact
Avoid any kind of external impact. External impacts can generate sparks through
friction processes between different materials.
16. Instrument filling
The filling/refilling of instruments leads to a loss of the explosion protection and
can lead to damage to the instrument.
14
WIKA operating instructions pressure gauge model 5 per ATEX

Publicité

loading