Memory module:
-
No values with unique ID are stored or printed:
→ Initialize memory in service menu (following the scales service manual).
-
The unique ID does not increment, and no values are stored or printed:
→ Initialize memory in menu (following the scales service manual).
-
Despite initialization no unique ID is stored:
→ Memory module is defective, contact service partner.
Real time clock module:
-
The time and date are stored or printed incorrectly:
→ Check time and date in menu (following the scales service manual).
-
The time and date are reset after disconnecting from the power supply:
→ Replace the button battery on of the real time clock.
-
Despite new battery date and time are reset when removing the power supply:
→ Real time clock is defective, contact service partner.