1
General information on the Alibi memory option .................................... 1
2
Description of the components................................................................. 2
3
Protection of stored legally relevant data and data loss prevention
measures ................................................................................................................... 2
4
Trouble shooting ........................................................................................ 2
1 General information on the Alibi memory option
-
For the transmission of weighing data provided by a verified scale via
an interface, KERN offers the alibi memory option YMM-03
-
This is a factory option, which is installed and preconfigured by
KERN, when a product containing this optional feature is purchased.
-
The Alibi memory offers the possibility to store up to 250.000
weighing results, when the memory is exhausted, already used IDs
are overwritten (starting with the first ID).
-
By pressing the Print key or by KCP remote control command "S" or
"MEMPRT" the storage process can be performed.
-
The weight value (N, G, T), date and time and a unique alibi ID are
stored.
-
When using a print option, the unique alibi ID is also printed for
identification purposes as well.
-
The stored data can be retrieved via the KCP command "MEMQID".
This can be used to query a specific single ID or a series of IDs.
-
Example:
o MEMQID 15 → The data record which is stored under ID 15 is
o MEMQID 15 20 → All data sets, which are stored from ID 15 to
TYMM-A-BA-def-2210
Alibi-Memory option including real time clock
module
Version 1.0 2022-12
Operating instructions
returned.
ID 20, are returned.