Manual Start; Alarm Visualisation; Measurement Stop - JRI SPY RF U Manuel D'utilisation

Masquer les pouces Voir aussi pour SPY RF U:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

g)

Manual start

Press shortly on the pushbutton
It displays the temperature in °C degrees, channel number, measurement unit and memory status.
The green LED flashes every minute.
h)

Alarm visualisation

The SPY RF is equipped with different alarm indicators, when a threshold limit is overpassed.
Pre alarm
Alarm
Threshold indicator
High or low
i)

Measurement stop

Depending on the configuration, the SPY RF can stop recording or not. The different options are:
Rolling memory: once the memory is full, the new values replace the old ones.
Full memory: the recorder stops when its memory is full.
With the software: you can put the SPY RF in standby mode with Sirius when you do not use your
recorder.
With the pushbutton: this option is valid only if the SPY RF is configured in transport mode with a start by
pushbutton.
©JRI
+
Value measured
Flashes every 15"
Working LED (green):
2'' => starting measurements
+
then flashes every 1 minute
Alarm LED (red):
Flashes every 15''.
To stop your SPY RF, press between 5
and 10" on the button:
The 2 LEDs are on and then flash
-
alternatively.
The screen goes off, Halt goes on.
-
19

Publicité

Chapitres

Table des Matières
loading

Table des Matières