Electromagnetic Compliance (EMC)

Warning:
Note important information on electromagnetic compliance (EMC). The growing
number of electronic devices like PCs, displays, TV-sets, washing machines and mobile phones
may subject medical devices to electromagnetic interference. This may cause malfunctioning of
the medical device and create potentially unsafe situations. Equally medical devices should not
interfere with other electronic devices. The norm IEC60601-1-2 was introduced to regulate the
requirements for EMC and prevent unsafe product situations. This norm defines the degrees of
immunity against electromagnetic interference and the maximum electromagnetic emission
levels for medical devices. CardioSecur PRO complies with norm IEC60601-1‐2:2001 concerning
immunity as well as emissions. Nonetheless special precautions should be taken: Do not use
devices that cause strong electromagnetic fields in the vicinity of CardioSecur PRO. A minimum
distance of 2 meters is recommended. In case that the distance is less than 2 meters CardioSe-
cur PRO may not perform properly. In accordance with the norm IEC60601-1‐2:2001 you may
request our customer support (see chapter 12 Contact below) for further documentation.
DIN EN 550112
Industrial, scientific and medical high frequency devices (ISM devices)
Radio frequency interference – limit values and measuring procedure
Interference emission
Interference voltage
RFI field strength (not accredited testing)
10
Limit value class
Result
--
B
n/a
OK
V6.0