Cautio
(1) To achieve an accurate reading, user should
ensure there is no alcohol residue inside the
chamber before starting a new alcohol test.
(2) After any 4 consecutive tests, please allow the
device to rest for 2 minutes.
(3) When the product is unused, please store it in a dry environment.
(4) When low battery indicator flashes, the reading is
invalid, lease replace the battery (alkaline AAA
battery) immediately.
(5) If the difference between two successive reading
is >0.1‰, user should test again.
(6) If there is very high concentration tested,
wait
forat least 3 min. to preform another test
Specification
Range
Power Input
Display
Auto Power Off
Breath
Sampling
Recycle Time
Dimension
Weight
Alert Level
Agency
Approval
Accessory
n
0 / 0.08 ~ 1.80 ‰ BAC / HI (0.008 ~0.180%BAC)
3 x 1.5V AAA Alkaline battery
15 seconds
Digital LCD display with bi-color Backlighting
≥5 seconds continuous breath sample
≈25 seconds
about 12cm (L) x 6.5cm (W) x2.4cm(T)
about 40g ( Device only )
0.2‰
0.3‰
0.8‰BAC(US/UK)
CE
6 pcs Blowing tubes
user should
0.5‰BAC