Télécharger Imprimer la page

Hameg Instruments HM7042-5 Manuel page 19

Publicité

Les langues disponibles

Les langues disponibles

with a safety ground connection or via isolation transformers
of class 2. The instrument may be used in any position,
however, sufficient ventilation must be assured as convection
cooling is used. For continuous operation prefer a horizontal
or slightly upward position using the feet.
Do not cover either the holes of the case nor the
cooling fins.
Specifications with tolerances are valid after a 30 minute
warm-up period and at 23 degrees C. Specifications without
tolerances are typical values of an average instrument.
STOP
Warranty and Repair
HAMEG instruments are subject to a strict quality control. All
instruments are burned in for 10 hrs prior to shipment. By
intermittent operation almost all early failures are detected.
After burn-in a thorough test of all functions and of quality is
run, all specifications and operating modes are checked.
In case of reclamations during the two years warranty period
please contact the dealer from whom you purchased your
HAMEG instrument. Customers from the Federal Republic of
Germany may directly contact HAMEG for warranty processing
in order to speed up the procedure
The proceeding of repairs during the warranty period is subject
to our terms of warranty which are available on our web-site
(http://www.hameg.com). Even after expiry of the warranty
period please do not hesitate to contact our HAMEG customer
service for repairs and spare parts.
Return Material Authorization (RMA):
Before sending back your instrument to HAMEG do
apply for a RMA number either by fax or on the
Internet: http://www.hameg.de.
If you do not have suitable packaging for the in-
strument on hand please contact the HAMAG sales
STOP
department (Tel.: +49 (0) 6182/800 300, E-mail:
vertrieb@hameg.de) to order an empty original
cardboard box.
Maintenance
The instrument does not require any maintenance. Dirt may
be removed by a soft moist cloth, if necessary adding a mild
detergent. (Water and 1 %.) Grease may be removed with
benzine (petrol ether). Displays and windows may only be
cleaned with a moist cloth.
Do not use alcohol, solvents or paste. Under no
circumstances any fluid should be allowed to get
into the instrument. If other cleaning fluids are
STOP
STOP
used damage to the lacquered or plastic surfaces
is possible.
Mains voltage
A main voltage of 115V and 230V can be chosen. Please check
whether the mains voltage used corresponds with the voltage
indicated by the mains voltage selector on the rear panel. If
not, the voltage has to be changed. In this case the line fuse
has to be changed, too.
Please note:
After changing the main voltage, the line fuse has
to be changed. Otherwise the instrument may be
destroyed.
All manuals and user guides at all-guides.com
I m p o r t a n t
Changing the line fuse
The fuses are accessible from the outside and contained in
the line voltage connector housing. Before changing a fuse
disconnect the instrument from the line, the line cord must
be removed. Check fuse holder and line cord for any damages.
Use a suitable screw driver of appr. 2 mm to depress the
plastic fuse holder releases on both sides, the housing is
marked where the screw driver should be applied. After its
release the fuse holder will come out by itself pushed forward
by springs. The fuses can then be exchanged, please take care
not to bend the contact springs. Reinsertion of the fuse hol-
der is only possible in one position and by pressing against
the springs until the locks engage.
It is forbidden to repair defective fuses or to bridge them by
any means. Any damage caused this way will void the warranty.
Types of fuses:
Size 5 x 20 mm; 250V~,
IEC 60127-2/5
EN 60127-2/5
Line voltage
230 V
115 V
h i n t s
22
23
Correct fuse type
2 x 2,5 A slow blow
2 x 5 A slow blow
19
Sous réserve de modifications

Publicité

loading