Cardinal Health Kangaroo ePump Guide De Dépannage

Publicité

Les langues disponibles

Les langues disponibles

Liens rapides

Flushing Mode Operation
Kangaroo
ePump
Flushing Mode Operation "Flushing Set Required" Directions For Use
Battery Operation
Unplugging the pump from the AC line will automatically put the pump on battery power.
To recharge the battery, plug the pump into an AC wall outlet. Pump will automatically begin charging.
General Instructions
1. Fill or spike feeding container and water bag prior to setting up pump.
2. Turn pump on - Press
Power
button in lower right-hand corner.
3.
In order to achieve proper accuracy, the bottom of the feeding set bag must be 18 inches above the
top of the feeding pump.
4. Select u
"Keep Settings"
or u
"Clear Settings"
the u
"Keep Settings"
option should be selected).
5. Load the set per diagram.
Do not overstretch
Priming the Pump
1. Press u
"Prime Pump"
to access the pump priming options.
2. Press u
"Auto Prime"
to automatically prime the pump set. The pump will quickly prime both feed and
flush lines starting with the flush line. The formula will stop before reaching the end of the feeding line.
3. Press the u
"Hold To Prime"
menu selections to manually top off the line.
4. Press u "Done".
Setting the Feed Rate
1. Select u
"Adjust Feed"
then u
"Feed
Rate". Use the buttons on the left to program the pump from 1 to
400 in increments of 1 ml. Select u
"Enter"
Note the Volume To Be Delivered "VTBD" Rate - is an Optional Feature
Only use the "VTBD" option if you want the pump to stop and alarm once a set amount of formula is delivered.
1. Select u
"Adjust Flush"
then u
"Flush Volume"
administered from 10 to 500 in increments of 1ml. Select u
2. Select u
"Flush Interval"
to define the time interval between the start of each flushing cycle from 1 to
24 hours in increments of 1 hour. Select u "Enter", Select u "Done".
3. Select u "Run". You will notice a small drop scroll down the screen when running.
Re-Priming the Pump After the Feed Bag Empties
1. A pump set bag that has been emptied will trigger the
set bag can be refilled to continue the feeding, but only after the pump set has been re-primed.
2. Disconnect the feeding line from the patient.
3. Refill the bag.
4. Press u
"Continue"
to begin running the pump.
5. Press u "Hold", then press u
"Adjust
Settings", then press u
6. Press u
"Hold To Prime Flush"
until the water has reached the valve in the pump.
7. Press u
"Hold To Prime Feed"
until formula reaches the stepped connector at the end of the set.
8. Press u "Done", then select u "Run".
To Change Rate or Clear Volume
1. Select u "Hold".
2. Select u
"Clear Vol"
to clear the volume.
3. Select u
"Adjust Settings"
to adjust all settings.
4. Select u
"Run"
to return to normal operations.
Pump sets should not be reused after 24 hours of initial usage.
Enteral Feeding Pump
(for day to day use with the same feeding regimen
tubing.
when desired rate is set.
to set the volume of water per flush cycle to be
"Enter"
when desired rate is set.
Feed Error
screen. In this condition the pump
"Prime
Pump".
Troubleshooting Guide
Listed below are some of the probable causes of alarm conditions and their corrections.
Menu Selections
Programming and Menu
Selection Buttons
Symptom
Probable Cause
Pump in
HOLDING
mode for more
Hold Error
than 10 minutes.
Occlusion in pump-patient line.
Flow Error
(downstream occlusion)
Empty bag or occlusion in
Feed Error
bag-pump line.
(upstream occlusion)
Flush Error
Empty bag or occlusion in flush line.
Flow Error
Pump Set in use beyond
Use > 24 Hrs
recommended length of time.
Pump Set
Pump Set not properly loaded in pump.
Dislodged
Battery Low
Battery charge too low.
Pump Set tubing not properly loaded
Rotor Error
on rotor, or some other unusual rotor
operating condition.
General error caused by many factors.
System Error
See Manual for error list.
15 Hampshire Street Mansfield, MA 02048
© 2018 Cardinal Health Canada. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO, Kangaroo, and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners. CA Lit. No. CA-PMR-0118, 18-121-E (11/12/2018)
Status LEDs
Red=Error
Yellow=Hold/Pause
Green=Normal Operation
Power Button
Corrections
Press u
CONTINUE
to return to
HOLDING
screen, or press and hold the
Power button
to turn off. Several options are available
from the
HOLDING
screen, such as
RUN
or
ADJUST
SETTINGS.
Check for and correct the occlusion in the
line between pump and patient. Replace
Pump Set if error cannot be resolved.
Check for empty bag and refill, or check for
and correct the occlusion in the line between
the pump and feed bag. Replace Pump Set if
error cannot be resolved.
Check for empty bag and refill, or check for
and correct the occlusion in the line between
the pump and flush bag. Replace Pump Set if
error cannot be resolved.
Informational warning message blinks on
upper left of screen. It is recommended that
the Pump Set be replaced.
Check black ring retainer (MISTIC) and
valve; reload Pump Set and restart pump to
continue. Replace Pump Set if error cannot
be resolved.
Immediately plug pump into A/C outlet to
recharge battery and to provide power for pump.
Check that the tubing is loaded on the rotor
and that it is not damaged, torn, etc. Reload
tubing or replace Pump Set.
Power down pump and attempt restart to
clear error. Call customer service with screen
code if error cannot be resolved.

Publicité

loading

Sommaire des Matières pour Cardinal Health Kangaroo ePump

  • Page 1 15 Hampshire Street Mansfield, MA 02048 © 2018 Cardinal Health Canada. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO, Kangaroo, and ESSENTIAL TO CARE are trademarks or registered trademarks of Cardinal Health. All other marks are the property of their respective owners. CA Lit. No. CA-PMR-0118, 18-121-E (11/12/2018)