6.2 Maintenance
(fuse replacement)

Check before each use

Visually inspect the device for signs of me-
chanical damage before each use.
If you detect damage or malfunctions that
may impair the safety of the patient or
of the operator, have the device repaired
before using it.
Technical inspections
Due to their design, the bearings and joints
of the ARTROMOT®-SP3 do not require
maintenance and all materials are protected
from corrosion, BUT: For safety, the devices
require regular maintenance. To maintain
the functional and operational safety, check
all components for damage and loose con-
nections at least once a year.
These checks should be performed by per-
sons with adequate training and experience.
Damaged and worn parts must immediately
be replaced with original spare parts by
authorized staff.
Warning!
m
Patient hazard, equipment malfunction
and damage –
– Refer repair and maintenance to
authorized persons. Persons are
authorized after training by a special-
ist trained and commissioned by the
manufacturer.
ORMED GmbH will make all docu-
ments required for servicing, such as
circuit diagrams, parts lists, descrip-
tions or calibration instructions,
available to authorized experts.
The inspections can be carried out by DJO
Technical Service within the framework of a
service agreement. Contact DJO for details.
Other than that, the manufacturer does
not require any other regular maintenance.
Note!
Regarding further technical or other
inspections or inspection intervals,
observe applicable local requirements,
such as IEC 62353, DGUV3 or similar
regulations to be observed by users of
medical equipment or electric devices.