VWR International VWR24086A Manuel D'instruction page 9

Table des Matières

Publicité

Section 1
Installation and Start-Up
1-2
VWR Collection
Certifications
Déclaration de conformité est disponible sur demande dans un document
distinct. Contactez VWR pour plus d'informations.
Safety Specifications
Indoor Use Only
Altitude - Up to 2,000 meters
Temperature - 15°C to 32°C
Humidity - Maximum RH 80% for temperatures up to 31°C, decreas-
ing linearly to 50% RH at 40°C
Mains Supply Fluctuations - Mains supply voltage fluctuations not to
exceed ±10% of the nominal voltage
Installation Category II
Pollution Degree 2
2
Class of Equipment I
1 Installation category (overvoltage category) defines the level of transient overvoltage which the
instrument is designed to withstand safely. It depends on the nature of the electricity supply and its
overvoltage protection means. For example, in CAT II which is the category used for instruments in
installations supplied from a supply comparable to public mains such as hospital and research
laboratories and most industrial laboratories, the expected transient overvoltage is 2500V for a
230V supply and 1500V for a 120V supply.
2
Pollution degree describes the amount of conductive pollution present in the operating
environment. Pollution degree 2 assumes that normally only non-conductive pollution such as dust
occurs with the exception of occasional conductivity caused by condensation.
1
VWR International