Z-Lock:
Utilized to prevent inadvertent re-zeroing of
the gauge.
1. In Menu mode, press
"ZLOCK" appears.
|
|
2. Press
3. Press
,
OFF" (default).
|
|
4. Press
will briefly show "DONE") and return to
Measurement mode.
Note: If Z-Lock is activated, the gauge will
display "ZLOCK" if zero is attempted.
,
to select.
to select "L ON" or "L
to select value (display
Re-Calibration:
Provides the user the ability to field calibrate
the product. Original factory calibration is
permanently retained in memory and can be
until
recalled at any time.
1. In Menu mode press
"RECAL" appears; proceed to press
|
|
2. Display will indicate "FACT" (factory)
"FIELD" or "NEW". Press
to scroll.
3. Pressing
is displayed will restore values of factory
calibration.
4. Pressing
displayed will restore values from latest
field calibration.
5. Pressing "Enter" key
"NEW" appears will enter the Recalibrate
mode.
6. When "NEW" is displayed, press
|
|
will flash CNFRM or CANCL. Press
or
"Enter" key
then flash CNFRM. Press ENTER,
display will flash "OK/ APPLY/ REF/
PSI/ THEN/ PRESS/ ENTER/ TO/ START/
OR/ OTHER/ TO/ ABORT". Numeric
display will read .00000. Vent sensor
to atmospheric pressure. Press
|
|
count down from 6 seconds then briefly
display CAL then automatically go to next
step.
7. Display flashes "APPLY/ REF/ PSI/ THEN/
PRESS/ ENTER/ TO/ START/ OR/ OTHER/
TO/ ABORT"; apply the full scale pressure
11
,
.
|
|
while "FACT" (factory)
|
|
when "FIELD" is
|
|
  ; upon release, the display
to select then press
|
|
  , the display will
, display will show "WAIT" and
until
,
when