Cleaning And Maintenance; Technical Characteristics - TUCKY 21 Notice D'utilisation

Masquer les pouces Voir aussi pour 21:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 8
Turn off Tucky
- To turn off Tucky, press the On/Off button; the LED will lit during two seconds. Otherwise, Tucky will
automatically turn off 30 minutes after being removed from the body.

Cleaning and maintenance

- After each use, remove the adhesive and clean Tucky with a cotton soaked with medical alcohol. Do not use
water, abrasive cleaners, or dishwashing products on Tucky. Not intended to be sterilized.
- To avoid battery issues, charge Tucky at least once a year.

TECHNICAL CHARACTERISTICS

Type of product
Thermometer references
Adhesives reference
Storage / transport conditions
Normal conditions of use
Display resolution
Displayed temperature range
Accuracy (maximum laboratory error
within the display temperature range)
Power supply
Charger cable
Charging voltage
Alerts and information
Bluetooth® technology
Protection class IP22
Supplied accessories
Lifespan
Required equipment
Compatibility
ELECTROMAGNETIC EMISSION TESTS
RF emission (CISPR 11)
Harmonic current emission
6
Direct mode clinical thermometer
- TUCKY-21, TUCKY-21W, TUCKY-21Y, TUCKY-21O, TUCKY-21P, TUCKY-21B & TUCKY-21N
- ADH-21-E1
For complete product (including adhesives) or for adhesives only:
Temperature: +10 to 27 °C (+50 to 80 °F); Relative humidity: 40 to 60 %.
For thermometer only:
Temperature: -20 to +55 °C (-4 to 131 °F); Relative humidity: 15 to 95 %
Temperature: +15 to 40 °C (59 to 104 °F); Relative humidity: 15 to 95 % (non-condensing)
Atmospheric pressure: 76 to 106 kPa
0,1 °C (0.2 °F)
+24 to 44 °C (75 to 111 °F)
±0.1 °C (0.2 °F) from 37 to 39 °C (98 to 102 °F);
±0.2 °C (0.3 °F) from 32 to 37 °C (90 to 98 °F) and from 39 to 42 °C (102 to 107 °F);
±0.3 °C (0.5 °F) beyond these ranges.
Rechargeable internal Lithium battery not accessible. Capacity: 35 mAh
USB 2.0 Type A male to Micro USB Type B male cable
5V
High temperature: >38.4 °C (101.1 °F) (modifiable),
Rollover on the belly (deactivated),
Battery discharged (<6 %), Tucky detached, Long connection loss (>15min), Error
BLE 4.2; very low output power (1.35 mW)
Protected against solid bodies larger than 12 mm
Protected against dripping water up to 15° from vertical
Instruction for use, charger cable, double-sided adhesive tape ADH-21
Thermometer: 5 years (500 charge cycles); Adhesives: 5 years
Smartphone or tablet with Bluetooth and internet connectivity. A list of recommended
equipment is available on the manufacturer's website
From iPhone 4S, iPad 3 and Android 4.3
The device is intended for use in the electromagnetic environment defined below
COMPLIANCE
Group 1
Class B
Not applicable
ENVIRONMENT
RF emissions are very low and are not likely to cause interference
to nearby electronic equipment
Conform
Battery powered