ThermoProbe TL3-A Mode D'emploi page 7

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 13
INTRINSIC SAFETY
Intrinsically safe equipment is defined as "equipment and wiring which is incapable of releasing sufficient electrical
or thermal energy under normal or abnormal conditions to cause ignition of a specific hazardous atmospheric mixture
in its most easily ignited concentration." (ISA-RP12.6) This is achieved by limiting the amount of power available to
the electrical equipment in the hazardous area to a level below that which will ignite the gases.
In order to have a fire or explosion, fuel, oxygen and a source of ignition must be present. An intrinsically safe system
assumes the fuel and oxygen are present in the atmosphere, but the system is designed so the electrical energy or
thermal energy of a particular instrument loop can never be great enough to cause ignition.
BATTERIES
Batteries must be changed in Non-hazardous area.
Batteries must be of correct approved type.
Batteries must not be mixed with batteries of other models or manufacturers.
Batteries must be installed with correct polarity making sure the (+) end of the battery is aligned with (+)
symbol embossed in the battery case.
CERTIFIED Batteries for the TL3-A, TL3-R and TL3-W are as follows:
Model
TL3-A
TL3-W, TL3-R
FCC Compliance Statement (Bluetooth Models Only)
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This
device may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.
CAUTION: The grantee is not responsible for any changes or modifications not expressly approved by the party
responsible for compliance. Such modifications could void the user's authority to operate the equipment.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in
a residential installation. This equipment generates, uses, and can radiate radio frequency energy, and if not installed
and used in accordance with the instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipment off and on, the user
is encouraged to try to correct the interference by one or more of the following measures:
– Reorient or relocate the receiving antenna.
– Increase the separation between the equipment and receiver.
– Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
– Consult the dealer or an experienced radio/TV technician for help.
This equipment has been tested and meets applicable limits for radio frequency (RF) exposure. This transmitter must
not be co-located or operating in conjunction with any other antenna or transmitter.
Type
CR2032 LiMnO
2
AAA (LR03) Alkaline
Duracell
DL2032
MN2400
Panasonic
CR2032
LR03XWA
07/2022, JK

Publicité

Table des Matières
loading

Produits Connexes pour ThermoProbe TL3-A

Ce manuel est également adapté pour:

Tl3-wTl3-r

Table des Matières