USER INSTRUCTIONS – TL3-A, TL3-W & TL3-R
INTRODUCTION
This manual describes the basic function, use and safety instructions for the model TL3-A, TL3-W and TL3-R portable
digital reference thermometer instruments.

REPLACING BATTERY

When the battery voltage is low, the display will show 'Lo bAtt'.
Replace the battery as soon as possible in a safe location after the low battery is noticed. This will avoid possible
malfunctioning. Do not attempt to calibrate the instrument if the low battery indicator is displayed.
WARNING:
Batteries must be changed in Non-hazardous area.
Batteries must be of correct approved type.
Batteries must be installed with correct polarity making sure the (+) end of the battery is aligned with (+)
symbol embossed in the battery case.
Batteries must not be installed with polarity reversed where one cell could charge another cell.
a) Ensure the instrument is in a non-hazardous area & powered off.
b) Use a #2 Phillips drive to remove the 1 screw holding the back cover on the TL3.
c) Replace each new battery making sure the (+) end of the battery is aligned with (+) symbol embossed in the
battery case.
d) Reinstall the cover. Make sure the rubber gasket is seated in its groove around the case opening (TL3-W, TL3-R).
CERTIFIED Batteries for the TL3-A, TL3-W and TL3-R are as follows:
Model
TL3-A
TL3-W, TL3-R
AUTHORIZED REPAIR
It is recommended that service beyond the scope of this manual be performed by ThermoProbe, Inc. or one of its
authorized distributors.
See
www.thermoprobe.net
Type
(1) CR2032 LiMnO
2
(2) AAA (LR03) Alkaline
for video on proper use of this instrument.
Duracell
DL2032
MN2400
Panasonic
CR2032
LR03XWA
07/2022, JK