Care And Maintenance - CAPTELEC BODY CHECK Manuel

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

T 1090 / YK-IRT1
Pressing down"Power"button for 3 seconds when the device in off state,the unit will be switched between ℃/℉.
Cautions:This operation is unable to recycle switch,turn off and operate as 3.3 again to switch the unit back.
6) Fever prompt function
Thermometer has fever prompt function. When body temperature is over 38.0℃ during measurement, the thermometer
will give out a"beep⸺beep⸺beep"sound in order to remind the person who is under test that he/she gets a fever. At
the same time, backlight of 3 different colors will indicate the state of temperature: normal, on the high side or fever.
Green backlight: Below 37.5 ℃, indicating temperature is normal; Yellow backlight: Between 37.6~38.0℃, indicating
temperature is on the high side(should pay attention to the temperature); Red backlight: over 38.1℃, indicating fever(
should see a doctor as soon as possible).
Sometimes parents tend to rely on their principal in demand on their child's forehead, it is even a reflex whether the child
is feverish or not. Or, it suffices that this hand is cold or hot to misjudge the temperature. With the forehead thermometer,
the results obtained are quite precise, provided you follow the correct procedure.
You will add 0.3 ° C to the evaluated value to get the exact internal temperature. In the case of a forehead measurement
the value to be added will be 0.2 ° C.
E. TROUBLESHOOTING
Phenomenons
Flicker
Blank screen
Lo
Hi
F. CARE AND MAINTENANCE
Clean the product only with a dry or slightly damp, lint-free cloth.
Possible Reasons
Power deficiency
Thermometer is power off
automatically
Battery isn' t installed properly
Battery has no power
Screen is still blank
TEMP is too low
Environment TEMP is too low
TEMP is too high
Environment TEMP is too high
Solutions
Replace battery immediately
Restart by pressing the power key
Check the battery board
Replace battery immediately
Contact distributor and send back the
product for repairation
⸻—
Measurement in proper environment
⸻—
Measurement in proper environment