4. Operation

4.6. Other options for calibration

Automatic buffers recognition
Technical buffers
pH 2.00, 4.01, 7.00, 9.21 and 10.00 at 25°C (77°F).
Buffers DIN 19266
pH 1.679, 4.006, 6.865, 7.000, 9.180, 10.012 and 12.454.
User's buffers
Suitable when the two above mentioned options will not be
used. The sensION
table pH/Temp. for the desired buffers. See System, page 16.
Calibration to X value
Manual re-adjustment of the measured pH at any value of the
scale. The pH readjustment is equivalent to calibration with only
one buffer.
10-04-2010
17:45
ÉTALONNAGE pH
Nouvel étalonnage
dans 1 jour
Tampons tecnics
...
Démarrer étalonnage
Start calibration
Important note:
Once the calibration is
re-programmed, it will be
performed with the new
programming.
See Stability Criteria
and View mV
on page 11.
sensION+ PH31 HACH_LANGE 0710.indd ANG:9
sensION+ PH31 HACH_LANGE 0710.indd ANG:9
All manuals and user guides at all-guides.com
+
allows introducing the
PH 31
Étalonnage pH
Étalonnage pH
C.stabilité:standard
Type d'étalonnage:
Type d'étalonnage
Tampons techniques
Frequence
Tampons DIN 19266
Visualisez mV
Tampons opérateur
A un valeur X
Étalonnage pH
Type d'étalonnage:
Tampons techniques
Tampons DIN 19266
Tampons opérateur
A un valeur X
Étalonnage pH
Type d'étalonnage:
Tampons techniques
Tampons DIN 19266
Tampons opérateur
A un valeur X
Étalonnage pH
Type d'étalonnage:
Tampons techniques
Tampons DIN 19266
Tampons opérateur
A un valeur X
Étalonnage pH
Type d'étalonnage:
Introduct. données
Étal. théorique
Étalonnage pH
Type d'étalonnage:
Introduct. données
Étal. théorique
Étalonnage pH
Étalonnage pH
C.stabilité:standard
Frequence étal.
Type d'étalonnage
Frequence
Avis automatique.
Visualisez mV
1 jour
00
heures
Calibration data introduction
Manual introduction of calibration data for an electrode cali-
brated previously, for example in calibration laboratory.
Theoretical calibration
Selecting "theoretical calibration" the calibration data of the elec-
trode in use are replaced by the theoretical values at 25°C (77°F).
This step is prior to the obtention of a calibration certificate.
Calibration frequency
Select between 0 h and 7 days. The instrument by default is
programmed for daily calibration and indicates the remaining
time to the new calibration.
Selecting 0 h the automatic re-calibration warning is deactivated.
Stability criteria.
Rapid criterion (variation < 0.02 pH in 6 s).
Standard criterion (variation < 0.01 pH in 6 s).
Strict criterion (variation < 0.002 pH in 6 s).
It is highly recommended to apply the same stability criteria to
measure and calibrate.
ÉTALONNANT pH
Tampons techniques
Utiliser 1er tampon
Continuer...
ÉTALONNANT pH
ÉTALONNANT pH
Tampons DIN 19266
Tampons DIN 19266
Utiliser 1er tampon
Tampon pH 7.000
Continuer...
Tampon pH 6.865
ÉTALONNANT pH
ÉTALONNANT pH
Tampons opérateur
Tampons opérateur
Démarrer étal.
1er tampon pH
N. tampons: 1
Tampon 1
Continuer...
See page 17.
ÉTALONNANT pH
ÉTALONNANT pH
A un valeure X
pH 3.08
Utiliser étalon
Continuer...
Étalonnage pH
Introduction données
Pot. asym.: 0 mV
Pente: 59.16mV/pH
Température: 25 °C
Garder changes
The instrument goes to
MEASURE menu. The
introduced calibration
data will be used.
Étalonnage pH
The instrument goes
Étal. théorique
to MEASURE menu.
Pot. asym.: 0 mV
The theoretical cali-
Pente: 59.16mV/pH
Température: 25 °C
bration will be used.
Continuer...
The instrument goes to the
option "CALIBRATE pH" and
it must be recalibrated.
Start of calibration
with automatic
recognition of the
selected buffers.
ÉTALONNANT pH
Régler valeur
25˚C
+03.09 pH
00:06
The instrument
goes to Measure.
9
8/7/10 16:45:01
8/7/10 16:45:01