Uline H-8862 Manuel page 3

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

c
REAL TIME CLOCK SETTING
When using thermometer for the first time, set time
functions. With thermometer off, press and hold the
power on/off button
mode.
SET THE TIME FORMAT
The device can display the time
in either an AM/PM (12-hour) or
24:00 (24-hour) format. With the
preferred time format on the
display, press the the power on/
off button
. The hour format
flashes automatically. (See Figure 2)
SET THE HOUR
Press and release the start button to
advance one hour until the correct hour
appears. (See Figure 3) After the hour is set,
press the power on/off button
figure flashes automatically.
SET THE MINUTE
Press and release the start button to
advance one minute until the correct minute
appears. After the minute is set, press the
power on/off button
, the year figure
flashes automatically. (See Figure 4)
SET THE YEAR
Press and release the start button to
advance one year until the correct year
appears. (See Figure 5) After the year is set,
press the power on/off button
figure appears.
SET THE MONTH
Press and release the start button to
advance one month until the correct month
appears. After the month is set, press the
power on/off button
, the date figure
flashes automatically. (See Figure 6)
SET THE DATE
Press and release the start button to
advance one day until the correct day
appears. (See Figure 7) After the day is set,
press the power on/off button
setting mode.
PAGE 3 OF 21
for three seconds to enter setting
Figure 2
, the minute
, the month
to exit the
OPERATION
TEMPERATURE-TAKING HINTS
To ensure that the reading always reflects body
temperature accurately, note the following factors that
may affect an accurate reading:
1.
Know each individual's normal temperature when
they are well. This is the only way to accurately
diagnose a fever. To determine normal temperature,
take multiple readings when healthy. Re-measure
with a standard digital thermometer for confirmation.
2. Users and the thermometer should be in the same
ambient temperature for at least 10 minutes before
taking a reading.
3. Users should not drink, eat or be physically active
such as bathing, showering, shampooing and hair
drying before/while taking the measurement.
4. Oils or cosmetics on the forehead may give a lower
temperature reading than the actual one. Remove
dirt from the forehead before taking a measurement.
5. Do not hold a hand on the forehead for any length
of time.
Figure 3
6. Do not take temperature over scar tissue, open sores
or abrasions.
7.
Do not use the thermometer on a perspiring or
sweating forehead.
8. Do not take a measurement while or immediately
Figure 4
after nursing a baby.
9. Thermometer is intended for home and clinical use.
10. Do not take temperatures near places that are very
hot, such as fireplaces and stoves.
11. Do not touch the probe window. The probe window
of the thermometer is the most delicate part of
Figure 5
the device. The accuracy of the reading may be
affected if the probe window is damaged or dirty.
12. If thermometer is stored in a significantly different
environment than testing location, place it in testing
location for approximately 30 minutes prior to use.
13. It is not intended for use in oxygen rich environments
Figure 6
or in the presence of flammable anesthetic mixture
with air, oxygen or nitrous oxide.
Figure 7
0920 IH-8862

Publicité

Table des Matières
loading

Table des Matières