Télécharger Imprimer la page

Therabath TB6 Manuel D'utilisation page 2

Bain de paraffine
Masquer les pouces Voir aussi pour TB6:

Publicité

Les langues disponibles

Les langues disponibles

• Use paraffin wax that is formulated for use in a paraffin warmer. Paraffin that has a higher
or lower melting point should not be used in this unit.
• Use an electrically grounded (3-pin) receptacle that conforms to the applicable local and
national electrical codes.
• Do not operate in the presence of flammable anesthetic mixtures, oxygen, or nitrous oxide.
• Do not operate the unit without the plastic grate on the bottom.
• Do not touch the bottom or sides of the metal tank, or a burn may result.
• Do not heat paraffin on a stove, hotplate, or in a microwave.
• Do use this appliance on a level, secure surface.
• Do not use any additives in the paraffin except as allowed by the manufacturer.
• Do not modify this equipment without authorization from the manufacturer.
Cautions
• Meets IEC/EN 60601-1-2 electromagnetic compatibility interference safety standard.
(Care must be taken when operating this equipment around other equipment. Potential
electromagnetic or other interference could occur to this or the other equipment. Minimize
this interference by not using other equipment in conjunction with it.)
• Do not remelt used paraffin, which may pick up particles or chemicals that may damage
the tank.
• Never use sharp objects to remove hardened paraffin from the unit. See cleaning
instructions.
• Do not leave an empty unit plugged in.
• Do not pour melted paraffin from the tank.
Technical Information
• Power, Model TB6: 100–125V, 0.87–1.3A, 50–60Hz
• Power, Model TB7: 220–240V, 0.55–0.65A, 50–60Hz
• Heater: TB6: 87–162 watts, TB7: 121-156 watts
• Operating temperature: 126º–134º F. (52.2º–56.7º C.)
• 14.3" L x 8.2" W x 7.5" H (36.32 cm L x 20.83 cm W x 19.1 cm H)
• Capacity: 9 lbs. (4.08 kg.)
• Melt time: 6-8 hours
• Duty cycle: continuous operation
• Degree of protection against moisture ingress: IPX0
• 220V unit must be used with a hospital grade cord of at least the following maximum
ratings: 10 A, 220 VAC.
Environmental Ratings
Temperature:
Relative Humidity:
Atmospheric Pressure:
Heating System and Pilot Light Operation
• The Therabath is equipped with 2 thermostats. The Control
Thermostat automatically heats and maintains the paraffin at 130
degrees F (54.4 C) with ± 4 degrees F variation. If the Control
Thermostat fails, the High Limit Thermostat automatically shuts
off the unit completely. Return to manufacturer for servicing.
• The pilot light (on end of unit) indicates the unit is heating, and is
lit only during the heating cycle. During normal operation, the pilot
light goes on and off at regular intervals. Most of the time it is off.
• If the unit is not completely melted after being plugged in for 12
hours, the unit has malfunctioned and should be returned to the
manufacturer for servicing.
Operating Instructions
1. Unpack the unit and place it on a flat, secure surface.
2. Open the included bags of paraffin and place the contents inside the tank.
Operating
65° to 85° F (18.3° to 29.4° C)
10% to 95%
70 to 160 kPa
• 2 / 12 •
NOTE: "Hospital Grade"
grounding reliability can
only be achieved when the
equipment is connected to
an equivalent receptacle
marked "Hospital Only" or
"Hospital Grade"
NOTE: Current and wattage
must be measured at
operating temperature.
Transport and Storage
-40° to 130° F (-40° to 54.4° C)
10% to 95%
70 to 160 kPa