Nuvo B2AV1000-1 Manuel page 15

Table des Matières

Publicité

CAUTION:
IMPORTANT: Operate table through all articulations prior to usage when table has been
stored for longer than four weeks,
Plug power cord into a properly grounded socket only.
When using a third party arm support, keep an angle greater than 15 degrees along the
direction of table length to avoid component damage during tabletop movements.
Do not articulate table into any angle that could hurt the patient.
Do not place this surgical table near other equipment that will generate electromagnetic
interference or other types of interference.
Before using high-frequency surgical equipments, cardiac defibrillators or cardiac
defibrillator-monitors, please refer to their instruction manual.
Use only NUVO's mattress pads in order to avoid risk of electrostatic shock.
When the power cord is damaged in any way, replace it immediately or else use the
table's internal electrical power source until it can be replaced.
The antistatic pathway is: Mattress Pad→Tabletop→Column→Base→Floor Lock
Pads→Floor. Therefore, the anti-static properties of the table depend upon the use of
the recommended NUVO mattress pads.
If the surgical drape hangs over the articulation points or the articulation covers (stainless
steel plates covering the joints), movement could cause damage to the table. Tuck the
drapes between the mattress pad and tabletop in order to prevent this.
Static electricity could damage the IC. Avoid any bodily contact with the electronic
systems. Use probes if necessary. Personnel must use a grounding device when
probing circuits or connectors, or else immediate electrical failure could result.
There is only one way to insert the connector of the hand pendant into its receptacle on
the table. The connector must be inserted straight into the receptacle without twisting it,
or else pins inside the receptacle will be bent and rendered unusable.
Since collision with the table joint covers may damage them, keep stretchers or other
heavy objects away from the covers.
Read the entire instructional manual before operating the table.
MISE EN GARDE:
 Seuls les nettoyants et désinfectants recommandés doivent être utilisés sur les tables
NUVO. N'utilisez pas d'alcool pour procéder au nettoyage de la table, car ses propriétés
1-7
B2AV1000 Series Surgical Table
IM-B2AV1000-E01

Publicité

Table des Matières
loading

Table des Matières