WARNING:
table.
FALLING HAZARD: When a patient is on the table, do not release floor locks.
TRIPPING HAZARD: Make sure to route the power cord so that it will not create a tripping
hazard.
SHOCK HAZARD: Monitor the electrically conductive properties of floor locks. Measure
their resistance level according to the requirements of NFPA 99 before placing table in
use. Subsequent routine testing should be performed at least once every three months.
PATIENT INJURY HAZARD: A potential hazard exists when the surgical table is used
together with high-frequency surgical equipment such as cardiac defibrillators and cardiac
defibrillator-monitors. In certain conditions the antistatic properties of the mattress pads
could become insufficient to prevent burns to the patient's skin. Care must be taken to
ensure that the insulation between the mattress pad and the patient's skin is maintained,
including monitoring the bed sheet for excessive moisture, which could compromise this
insulation. Carefully read the manufacturers' instructions of the aforementioned
products, as they relate to patient safety.
ELECTRIC SHOCK HAZARD: Only authorized service personnel should be allowed to
remove covers.
INSTABILITY HAZARD: Using NUVO table and accessories or accessories
manufactured and sold by other companies for other than their stated purpose may result
in possible patient or user injury, table or accessory damage.
UNANTICIPATED TABLE MOVEMENT HAZARD: Secure patient to table according to
recommended positioning practices.
EXPLOSION HAZARD: Do not use table with flammable anesthetics.
INFECTION HAZARD: When cleaning the surgical table, personnel should wear gloves, a
mask, eye protection, and other safety equipment which will protect body parts from
exposure to aerosols which could be deflected from contaminated surfaces.
PERSONAL INJURY HAZARD: When installing any surgical table accessory, check for
correct attachment and tighten securely (if applicable). Do not use worn or damaged
accessories.
PERSONAL INJURY HAZARD: Only recommended cleaners/disinfectants should be
used on NUVO surgical tables. Do not use alcohol, which does not have sufficient
cleaning/disinfecting properties, or phenolics, which when inadequately rinsed off could
1-2
B2AV1000 Series Surgical Table
IM-B2AV1000-E01