Fault Screens
Door Fault: Displays when lamp door is not fully closed.
o
o
Fan Fault: Displays when fan is not operating correctly.
o
o
o
Lamp Fault: Displays when lamp is not functioning.
o
o
o
Dimmer Fault: Displays when dimming mechanism is not functioning.
o
o
LIT-263 Sunoptic Technologies
Rev. A
Close lamp compartment door.
Device will need servicing if error persists.
Turn device OFF.
Inspect fan area and ensure nothing is obstructing the fan.
Device will need servicing if error persists.
Turn device OFF.
Ensure lamp is correctly installed or install different lamp.
Device will need servicing if error persists.
Restart device.
Device will need servicing if error persists.
®
Page 12 of 134