Troubleshooting

Review the information in the table below to troubleshoot operating problems.
Problem on
display
Er1
Temperature probe fault, electronic board
fault or incorrect setting of parameters.
Er2
Momentary temperature in the incubator is at
least 5 ºC higher than SET temperature,
because the incubator has been working at
higher SET temperature recently.
Momentary temperature in the incubator is at
least 5 ºC higher than SET temperature, due
to temperature probe fault, electronic board
fault or incorrect setting of parameters.
Er3
Momentary temperature does not reach SET
temperature in 2 hours, because the door is
opened, you open the door too frequently or
the environment temperature is too low.
Momentary temperature does not reach SET
temperature in 2 hours, due to thermal fuse
cut off, heater fault, electronic board fault or
incorrect setting of parameters.

Cleaning and maintenance

Prior to any cleaning or maintenance of the unit, disconnect the power cord from the wall
outlet.
No routine maintenance is required on the unit.
Clean the unit internally and externally with a damp cloth and mild soap. Do not immerse the unit or spill
liquids into or on the unit. Use of alcohol or ammonia-based cleaners on the window may cause the window
to crack. Allow the incubator to dry completely inside and out before reconnecting the power and using.

Accessories

IL 10:
Description
Extra shelf, 226 x 215 mm, stainless steel
Tube rack, 2 x 6 positions, Ø 32 mm
IL 23:
Description
Extra shelf, 286 x 285 mm, stainless steel
Tube rack, 2 x 6 positions, Ø 32 mm
IL 56:
Description
Extra shelf, 385 x 345 mm, stainless steel
Tube rack, 2 x 6 positions, Ø 32 mm
EN 8
Cause
Quantity
1
1
Quantity
1
1
Quantity
1
1
Solution
Contact your distributor.
Check if SET temperature is for 5 ºC, or
more, lower than momentary temperature
in the incubator. If so, allow incubator to
cool down. Turn the power switch off and
on, then try again.
Contact your distributor.
Check that the door is securely closed,
don't open the door too frequently during
operation, check that environment
temperature is above minimum. Turn the
power switch off and on, then try again.