Instructions for use

Getting Started

Check the data plate for voltage, cycle, wattage and ampere requirements. If matched to your power source,
plug the power cord into a grounded outlet. Voltage should not vary more than 10% from the data plate
rating.
In selecting a location, consider all conditions which might affect performance, such as heat from radiators,
autoclaves, etc. Avoid direct sun, fast-moving air currents, heating/cooling ducts and high traffic areas. Allow
a minimum of 5 cm between the unit and walls or partitions which might obstruct free airflow.

Operation

Do not use this incubator in explosive or flammable environments. Do not heat or incubate
flammable, explosive or highly reactive materials in this incubator as serious injury may
result.
Place the unit on a level surface and plug the unit into a properly grounded outlet of appropriate voltage.
Install the shelf at the desired height, with the shelf edges up. If you intend to place the samples on the
bottom of the chamber, install the shelf so that it sits with the edges down on the chamber bottom, thus
creating an air gap between the chamber bottom and your samples.
Place the samples in the chamber and close the door. Make sure the door is securely closed.
Turn the power switch to the ON position. The switch should illuminate. On display all "888" are displayed at
first, then automatically detected power supply frequency "F50" or "F60", then software version "VX.X", and
finally the current temperature in the chamber. Incubator starts to heat.
Set desired temperature with UP or DOWN key. You must hold UP or DOWN key more than 2 sec. DOT on
DISPLAY blink. When DOT start blink (dot blink 3 sec) set temperature on desired value between room
temperature +5 ºC to +70 ºC. When desired temperature is set, display will after 3 sec automatically show
momentary temperature.
You can always look at SET temperature when hold UP or DOWN key more than 2 sec.

Temperature re-adjustment

Incubator has software for temperature re-adjustment. You must first measure temperature in middle of
chamber or in sample for minimum 2 hours. You must measure temperature with a certificated reference
temperature measuring instrument with precision 0.1 ºC or more. After 2 hours or more read temperature on
thermometer and compare it with temperature on LED display. Difference from thermometer and display is
value, which you will enter in software.
Example 1: Reference temperature is 37.9 ºC, temperature on display is 37 ºC. Difference is 37.9 - 37 = 0.9
Value is 0.9 (value which you enter in software).
Example 2: Reference temperature is 36.2 ºC, temperature on display is 37 ºC. Difference is 36.2 - 37 = -0.8
Value is - 0.8 (value which you enter in software).
Procedure for temperature re-adjustment:
Press UP and DOWN key at the same time for more than 5 seconds - on display appears "COr" (correction)
and DOT blink.
With UP or DOWN key define value (when dot blink). Pay attention on sign of value (plus or minus). After 3
sec when you will define temperature display will automatically show temporary temperature.
EN 7