Télécharger Imprimer la page

Beta 1760/2TM Mode D'emploi page 3

Publicité

Les langues disponibles

Les langues disponibles

ELECTRICAL SPECIFICATIONS
Temperature Scale:
Measurement Range:
Resolution:
Accuracy:
± 2°C
± 4°F
±(0.5%rdg+1°C)
± (0.8% rdg+1°C)
± (0.5% rdg+2°F)
Temperature Coefficient:
Input protection:
Reading Rate:
Input Connector:
Environmental
Ambient Operating Range: 0°C to 50°C (32°F to122°F)
Storage Temperature:
Relative Humidity:
Celsius (°C) or Fahrenheit (T) user-selectable
-50 °C to 1300°C -58°F to 2000°F
1 °C or 1°F, 0.1°C or O.1°F
Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.
-50 °C to 0°C
-58°F to 32°F
0°C to 1000°C
1000°C to 1300°C
32°F to 2000°F
0.1 times the applicable accuracy specification per °C
from 0°C to 18°C and 28°C to 50°C
(32°F to 64°F and 82°F to 122°F).
60V dc or 24V rms ac maximum input voltage on any combination
of input pins.
2.5 times per second.
Accepts standard miniature thermocouple connectors (flat blades
spaced 7.9mm, center to centier).
-20°C to 60°C (-4°F to140°F)
0% to 80% (0°C to 35 °C) (32°F tc 95°F)
0% to 70% (35°C to50°C) (95T to122°F)
3

Publicité

loading