Operating instructions

1.0 Introduction

The Thermo Scientific
chromatography (GC) systems. It detects minute leaks of any gas with a thermal
conductivity different from air. The reference gas inlet (Figure 1) draws in ambient air for
comparison to air drawn into the sample probe. A leak is indicated by both an LED light
display and an audible alarm.
If this instrument is used in any manner other than described in the manual, the CE
declaration is void.
Unit State -
Powered
Unit Battery
State
Power Button
Depressed
Unit Charging
State
Unit Operation
4
leak detector is specifically designed for use with gas
Leak Detector LED Indicator Display and Interpretation
LED Color
On/Off
On
Blue >
Red & Yellow
On
Blue
On
Blue
NONE
On/Off
Green
On/Off
Green
On/Off
Green
On
Red or Yellow
On
Red & Yellow
LED State
On/Blue flashes
once, red and yellow
cycle for 15 seconds/
Off
On/Steady
On/Constant
Flashing
NONE
On/Flashing
On/Steady
Off
On/Steady
On/Cycling for
4 seconds/Off
Interpretation
Unit is powering on and running
startup sequence
Ready for use, battery does not
require charging
IMPORTANT: Unit must be
charged
Battery has fully discharged,
charge immediately
Unit is plugged in to the AC adaptor
or USB cable and is charging a very
low battery (trickle charging)
Unit is plugged in to AC adaptor
or USB cable and is in full charge
mode
Unit is plugged into AC adaptor
or USB cable and the unit is fully
charged
Unit is indicating a difference in
thermal conductivity between probe
and reference
Zero set button has been depressed
and unit is rezeroing