Intended Use

The disk for temperature adjustment is used for
calibrating and adjusting the temperature of Sartorius
moisture analyzers MA50 or MA100. Equipment that
has been calibrated and adjusted with the temperature
adjustment disk can be used as reference devices or as
inspection, measurement and testing equipment. The
printed-out calibration record can be used as proof of
compliance with the standards for inspection and
measuring equipment. Every temperature adjustment
disk has a serial number (ref. ID no.) and a factory
calibration certificate. The quality of the temperature
adjustment disk should be similarly tested. We
recommend that you have the temperature adjustment
disk tested once a year. For this purpose, you must
send the temperature adjustment set back to Sartorius.
The moisture analyzer should be adjusted.
– at room temperature (approx. 20°C)
– under the same ambient conditions
– at regular intervals defined by the user
– using the same complete temperature adjustment set.
The moisture analyzer should be set up on a clean
workbench in surroundings that are supplied with
clean air. A change in any of the following ambient
conditions will affect the results obtained:
– Sunlight
– Drafts caused by open windows or doors
– Air conditioning or fume hoods
– Gases or dust particles.
Leave a space of at least 20 cm around the moisture
analyzer and any other objects in the direct vicinity of
it, and at least 1 m above the moisture analyzer. After
you start the analyzer, it will carry out calibration and
adjustment fully automatically.
Important Note:
Read the operating instructions carefully before
calibrating and adjusting the temperature.