Automatic Mode - HEBU medical HB 8881 Mode D'emploi Et De Service

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

HEBU Vacuum Cleaner
Commissioning:
➢ Adapt the suction hose
➢ Connect the mains plug.
➢ Switch on the device and start up

4.2 Automatic mode

➢ Insert the mains plug of the
HEBU oscillating saw at the
vacuum cleaner (1).
➢ Switch on the vacuum cleaner (2).
The lamp lights up and the vacuum cleaner
runs for around 0.5 seconds
as a function check. The vacuum cleaner
then goes to the standby mode (3)
Note:
The vacuum cleaners can be used together with a oscillating HEBU cast saw as well as the autopsy
model.
The vacuum cleaner is automatically switched on and off in the standby mode by the HEBU
oscillating saw, provided you have performed the following points beforehand:
-
Every time the vacuum cleaner is first switched on, it goes to the standby mode.
-
or after being switched on again at least 6 secs. after last being switched off.
♦ The vacuum cleaner has a run-on time of up to 6 secs. This is intended to ensure that the vacuum
cleaner hose is completely emptied.
♦ If the vacuum cleaner is switched on again less than 4 seconds after being switched off,
the suction process operates without automatic function. However the HEBU oscillating saw
can still be switched on and off.
♦ The voltage specification of the vacuum cleaner must agree with the voltage specification
of the HEBU oscillating saw.
3
EN
2
1
23

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Hb 8881v

Table des Matières