2

Specifications and Environmental Conditions

2.1

Specifications

Table 2 Base Unit Specifications
Base Unit Specifications
Temperature Range at
23 °C
Display Accuracy
Stability
Axial Uniformity at 40
mm (1.6 in)
Axial Uniformity at 60
mm (2.4 in)
Radial Uniformity
Loading Effect (with
a 6.35 mm reference
probe and three 6.35
mm probes)
Loading Effect (versus
display with 6.35 mm
probes)
Hysteresis
Operating Conditions
Environmental
conditions for all
specifications except
temperature range
Immersion (Well)
Depth
Insert OD
Heating Time
Cooling Time
Resolution
Display
Key Pad
Size (H x W x D)
Specifications and Environmental Conditions
9142
–25 °C to 150 °C
(-13 °F to 302 °F)
± 0.2 °C Full Range
± 0.01 °C Full Range
± 0.05 °C Full Range
± 0.07 °C Full Range
± 0.01 °C Full Range
± 0.006 °C Full Range
± 0.08 °C Full Range
0.025 °C
0 °C to 50 °C, 0 % to 90 % RH (non-condensing)
30 mm (1.18 in)
16 min: 23 °C to 140 °C
23 min: 23 °C to 150 °C
25 min: –25 °C to 150 °C
15 min: 23 °C to –25 °C
25 min: 150 °C to –23 °C
14 min: 350 °C to 100 °C
LCD, °C or °F user-selectable
Arrows, Menu, Enter, Exit, 4 soft keys
290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)
9143
33 °C to 350 °C
(91 °F to 662 °F)
± 0.2 °C Full Range
± 0.02 °C at 33 °C
± 0.02 °C at 200 °C
± 0.03 °C at 350 °C
± 0.04 °C at 33 °C
± 0.1 °C at 200 °C
± 0.2 °C at 350 °C
± 0.04 °C at 33 °C
± 0.2 °C at 200 °C
± 0.25 °C at 350 °C
± 0.01 °C at 33 °C
± 0.015 °C at 200 °C
± 0.02 °C at 350 °C
± 0.015 °C Full Range
± 0.2 °C Full Range
0.03 °C
13 °C to 33 °C
150 mm (5.9 in)
25.3 mm (1.00 in)
5 min: 33 °C to 350 °C
32 min: 350 °C to 33 °C
0.01 °
Specifications
9144
50 °C to 660 °C
(122 °F to 1220 °F)
± 0.35 °C at 50 °C
± 0.35 °C at 420 °C
± 0.5 °C at 660 °C
± 0.03 °C at 50 °C
± 0.05 °C at 420 °C
± 0.05 °C at 660 °C
± 0.05 °C at 50 °C
± 0.35 °C at 420 °C
± 0.5 °C at 660 °C
± 0.1 °C at 50 °C
± 0.6 °C at 420 °C
± 0.8 °C at 660 °C
± 0.02 °C at 50 °C
± 0.05 °C at 420 °C
± 0.1 °C at 660 °C
± 0.015 °C at 50 °C
± 0.025 °C at 420 °C
± 0.035 °C at 660 °C
± 0.1 °C at 50 °C
± 0.2 °C at 420 °C
± 0.2 °C at 660 °C
0.1 °C
24.4 mm (0.96 in)
15 min: 50 °C to 660 °C
35 min: 660 °C to 50 °C
25 min: 660 °C to 100 °C
13