Guidance and Manufacturer's Declaration – Emissions All Equipment and Systems
®
The
Belmont
Rapid Infuser, FMS2000
customer or user of the
Belmont
Emissions Test
Compliance
RF Emissions
Group 1, Class A
CISPR 11
Harmonics IEC
Complies or Not
61000-3-2
applicable
Flicker
Complies or Not
IEC 61000-3-3
applicable
Guidance and Manufacturer's Declaration—Immunity All Equipment and Systems
®
The
Belmont
Rapid Infuser, FMS2000
customer or user of the
Belmont
Immunity Test
IEC 60601 Test Level
Electrostatic
±6kV contact
Discharge (ESD)
IEC 61000-4-2
±8kV air
Radiated RF
3 V/m
IEC 61000-4-3
Electrical Fast
±2kV on AC Mains
Transient/burst
IEC 61000-4-4
Surge
±1kV Differential
IEC 61000-4-5
±2kV Common
Conducted RF
3 Vrms from
IEC 61000-4-6
150 kHz to 80 MHz
Power Frequency
3A/m
50/60Hz Magnetic
Field
IEC 61000-4-8
Voltage dips, short
>95% Dip for 0.5 Cycle
interruptions and
voltage variations
60% Dip for 5 Cycles
on power supply
input lines
30% Dip for 25 Cycles
IEC 61000-4-11
>95% Dip for5 Seconds
Chapitre 4 : Réglage des paramètres et entretien préventif
Table 201
is intended for use in the electromagnetic environment specified below. The
®
should assure that it is used in such an environment.
Rapid Infuser, FMS2000
Electromagnetic Enforcement – guidance
®
The
Belmont
Rapid Infuser, FMS2000
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Complies
Complies
Table 202
is intended for use in the electromagnetic environment specified below. The
®
should assure that it is used in such an environment.
Rapid Infuser, FMS2000
Compliance Level
±6kV contact
±8kV air
3 V/m
±2kV on AC Mains
±1kV Differential
±2kV Common
V1 = 3 Vrms
3A/m
>95% Dip for 0.5 Cycle
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5 Seconds
-43-
uses RF energy only for its internal
Electromagnetic Environment -
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are synthetic, the
relative humidity should be at least
30%.
Radiated RF frequency is negligible at
any single frequency
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
If interference occurs, it may be
necessary to position the
Belmont
further from
Rapid Infuser, FMS2000
sources of power frequency magnetic
field.
Power frequency magnetic fields
should be that of a typical location in a
typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
®
Belmont
Rapid Infuser, FMS2000
requires continued operation during
power mains interruptions, it is
recommended that the
Belmont
Rapid Infuser, FMS2000
be powered
from an uninterruptible power supply
or battery.
®
®