Specifi Cations - Hameg Instruments HZ56-2 Manuel

La pince ampèremétrique
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Specifi cations
Electrical Characteristics
(all accuracies stated at 23°C ±1°C)
Current range
Measuring range
Overload capacity
Output sensitivity
Accuracy (at 25 °C)
Resolution
Typical output noise level
Load impedance
Conductor position sensitivity ±1 % relative to center reading
Frequency range
di / dt following
Response time
Temperature coeffi cient
Operating humidity
Altitude
Power supply
Working voltage
(see Safety Standards)
Battery life
General Characteristics
Maximum conductor size
Output cable/connections
Output zero
Operating temp. range
Storage temperature range
(with battery removed)
Weight
20 A
rms or DC
AC
± 30 A
500 A
100 mV/A
±1 % of reading ± 2 mA
±1 mA
200 μV rms
> 100 kΩ 100 pF
DC to 100 kHz (0.5 dB)
20 A / μs
better than 1 μs
± 0.01 % of reading / °C
15 to 85 % (non condensing) C
2000 m
9 V Alkaline,
NEDA 1604/PP3 IEC 6LR61
300 V
rms or DC
AC
30 hours, low battery indicator
19 mm diameter
2 m cable terminated with a
BNC connector (50 Ohms)
supplied with a 4 mm safety
adaptor
Manual adjust via thumbwheel
0 to +50 °C
-20 to +85 °C
290 g
Subject to change without notice
15

Publicité

Table des Matières
loading

Table des Matières