Scaler handpieces and their accessories (nosepiece, LED ring, optical guide, cord pass-through depending on model) must be cleaned
with extreme care.
Do not immerse or place scaler handpieces in an ultrasonic cleaner.
Only the Nosepiece disassembled from Ultrawave™ or Ultrawave™ XS LED handpieces can be immersed and treated using the ultrasonic
procedure.
If the surface of the devices is damaged, ultrasonic cleaning may cause further damage.
It is the responsibility of the end user to ensure that all equipment used to treat Ultradent products is properly installed, validated,
maintained and calibrated.
Whenever possible, a washer/disinfector should be used for Handpieces and their accessories. Prevent the overloading of wash baskets
for ultrasonic cleaning or cleaning in a washer/disinfector.
Cleaning/sterilization cycle limits
Repeated treatment cycles including manual washing and sterilization have minimal effects on Ultrawave™ LED handpiece scaler
handpieces, LED ring and Optical guide.
Repeated treatment cycles that include ultrasonic cleaning, manual washing and sterilization have minimal effects on scaler Handpiece
Nosepieces.
End of service life is normally determined by wear and damage due to use.
Important
− Throughout the procedure, wipe blood and/or debris from the Handpiece to prevent it drying on the surfaces.
− In case of external irrigation tubing or configuration of the device with tank, after each use, rinse the irrigation tubing, Handpiece and
tip for 1 minute with an enzymatic or alkaline cleaning solution to prevent debris and contamination from drying on the device.
Rinse with distilled or purified water for 1 minute to remove the enzymatic or alkaline cleaner. Flush the irrigation tubing to remove
liquid from the inside.
J12949 • V1 • (13) • 03/2014 • RB62US020A