Sterilisation cycle

1) Press the 'SET' button for 3 seconds to start the cycle. Actual temperature will be shown in the
upper display (white), set temperature will be shown in the lower display (red)
2) When the sterilisation temperature has been reached, the time left to reach end of the cycle is
shown on the lower display.
3) The end of the sterilisation cycle is indicated by an acoustic alarm for few seconds.
4) The cycle can be stopped at any time by simply pressing SET button for five seconds
5) At the end of vessel cooling, open the lid to take out the sterilised material.

Panel commands

SET
▼▲
FNC
NOTES:
All programming operations must be performed in quick sequence otherwise the programming
device zeros. The programme can be stopped by pressing the 'SET' button for more than five
seconds. The only parameters that need to be selected are (SP1) temperature and (t in 1) time.
The selected parameters remain memorised even after the instrument is turned off.

Cleaning after use

1) The outside surface may be kept bright and shiny by cleaning with a good silver polish or simply
by washing with soap and water. Iron and various minerals in water may darken the inside but
this discoloration will in no way affect sterilisation. These stains are harmless and may be
removed by using a solution of water and cream of tartar. For each quart of water, use one
tablespoon of cream of tartar. Pour enough solution into the vessel to cover the discoloration (do
not fill over 2/3 full), then close cover securely. Place regulator on vent pipe and heat until 15
pounds pressure is reached. Switch off and allow the unit to stand for two to three hours.
Remove regulator, open and empty contents. Scour thoroughly with a soap impregnated steel
wool cleaning pad; wash, rinse and dry.
2) Each time the vessel is washed, remove the sealing ring and wash in warm, soapy water, rinse,
dry and replace in cover.
To start/stop the programme and
change parameters to be set.