4.5 WATER CONNECTION

The water used should have the following properties:
pH:
Impurities:
Chlorides:
Cl:
Recommended filters:
A) Fine filter.
If the water contains impurities such as sand, iron particles or floating substances, we recommend the use of
a fine filter at the water input.
B) Activated carbon filter.
If the water has a high chlorine content over 0.2 mg/l (ppm) (this information can be obtained from the
relevant water board), an activated carbon filter should be installed.
C) Installation of osmosis recirculation.
When the chloride concentration is above 150 mg/l (ppm) (this information can be obtained from the relevant
water board), an osmosis recirculation installation should be mounted. In this case, please remember that
the minimum conductivity value is 400 µS/cm.
D) Water descaling:
For water with a high level of limescale (without chloride load) the water should be treated. Systems: H+.
Interchange of ions or Kleensteam. We strongly advise against the use of sodium exchangers (normally
used in dish washers) due to the formation of sodium sediment and the delay in boiling with common salt.
For the selection of the filter systems (A, B, C, D), it is recommended the manufacturer BRITTA.

4.6 WASTE WATER CONNECTION

Water inlet hose Fig:12
Water hardness:
6,5 - 7,5
Ø < 0,08 mm
máx. 150 mg/l
0,2 - 0,5 mg/l
Before connection, look at the label that
shows which is the water inlet.
An inappropriate installation of the oven may lead to an undesirable operation of the
device.
Therefore you must install the drain KIT supplied by the manufacturer (Fig. 12). This
drain must preferably be connected to a grid or an open tank.
The installation (Fig. 13 - Fig.14) should be done so that the installed drain outlet is
below the outlet of the oven with one enough slope to ensure its proper drainage (5 or
3 °).
Ensure that the holes of the rise tube are directed towards the opposite side of the
back panel to avoid condensation.
Observe the dimensions for the drain:
Pumping volume of the steam generator in a small amount of time: 0.7 l
sec.
Average residual water temperature: 65 ° C.
Only connect drinking water to the appliance.
Connect the appliance to the mains water
supply at point 1 using the hose supplied.
The pressure of the incoming water should
be between 200 and 400 kPa (2-4 kg/mm
We recommend 250 kPa
.
5 - 10 ºfH
7 - 14 ºeH
9 - 18 ºdH
400 – 1.000 µS/cm
Conductivity:
2
).
((French degrees)
(English degrees)
(German degrees)
53