BTL7-A/C/E/G5 _ _ -M _ _ _ _ -HB/WB-KA_ _/K_ _/FA_ _/F_ _
Micropulse Transducer - Rod Style
6
Calibration procedure (continued)
6.3.2

Adjusting

The detailed procedure for adjusting is
described on page 17 ff.
The factory-set null point and end point are replaced by a
new start point and end point and the associated output
values can be adjusted. The start and end values can be
adjusted as you like up to the limits.
Steps
► Move magnet to the new start position.
► Read new start point by activating the programming
inputs.
► Set the new start value by activating the programming
inputs.
New start value
New null point
Fig. 6-4:
Adjust a new start value
www.balluff.com
► Move magnet to the new end position.
► Read new end point by activating the programming
inputs.
► Set the new end value by activating the programming
inputs.
New end value
Before
Fig. 6-5:
Adjust a new end value
After
6.3.3

Reset

Restoring the transducer to its factory settings.
The detailed procedure for the reset is
described on page 19.
Before
After
New end point
english