Reply to control command «readsetup»
2 whole numbers are returned as a reply;
Code1_Code2. These must be converted into
binary format for decoding.
Code1
msb
#### #### #### ####
|||| |||| |||| |***-Pressure unit
|||| |||| |||| *----Temperature unit
|||| |||| |||*------Humidity unit
|||| |||| ||*-------Altitude unit
|||| |||| |*--------Time unit
|||| |***-*---------Storage interval
|||* *--------------Baudrate
***-----------------Cut-out time
Code2
msb
#### #### #### ####
|||| |||| |||| ||**-Mixed Mode
**** **** **** **---don't care
Code-decoding of contr. command «readsetup»
Code
Configuration
Pressure unit
Pressure unit
Pressure unit
Pressure unit
Pressure unit
010 hPa
011 mmHg
100 inH2O
101 inHg
110 psia
111 mbar
Temperature unit
Temperature unit
Temperature unit
Temperature unit
Temperature unit
0 °F
1 °C
Humidity unit
Humidity unit
Humidity unit
Humidity unit
Humidity unit
0 %rH
1 %rF
Altitude unit
Altitude unit
Altitude unit
Altitude unit
Altitude unit
0 ft
1 m
Time unit
Time unit
Time unit
Time unit
Time unit
0 /hours
1 /minutes
Cut-out time
Cut-out time
Cut-out time
Cut-out time
Cut-out time
011 30 min.
100 60 min.
101 continous
110 1 min.
111 10 min.
Error messages for control commands
er_00
Syntax invalid
er_01
False argument
er_02
Command does not coincide with
configuration
er_03
Remote-command incorrect
24
lsb (binary)
lsb (binary)
Code Configuration
Storage interval
Storage interval
Storage interval
Storage interval
Storage interval
0000
10 s
0001
20 s
0010
30 s
0011
1 min
0100
2 min
0101
5 min
0110
10 min
0111
20 min
1000
30 min
1001
1 h
1010
3 h
1011
6 h
1100
24 h
1101
manuell
1110
1 s
1111
5 s
Baudrate
Baudrate
Baudrate
Baudrate
Baudrate
00
1200 Baud
01
2400 Baud
10
4800 Baud
11
9600 Baud
Mixed Mode
Mixed Mode
Mixed Mode
Mixed Mode
Mixed Mode
01
QNH
10
Altitude
11
Baro