Problem./.Message
-- Load/Unload Weight --
Unit change not allowed !

-- Out of range --

8
Recommended Settings and Printer Models
Connected.Device.
AB-S/FACT, AB-S, AB-L,
PB-S/FACT, PB-S, PB-L,
AL, PL, PL-S, PL-L,
JB-G/FACT, JB-C/FACT,
JB-L-G, JB-L-C,
JL-G,JL-C, JL-G/L, EL,
AB-S/PH, PB-S/PH
PG-S, PG-S/PH, SB
XP, XS
Katakana
Russian
Chinese,
Japanese
MJ33, HB-S43
Nihongo
Russian
Possible.cause
• No sample or the same
• Changed unit during
• Sample weight is out of
Device.Settings:
Printer Mode
Printer Mode
Printer Mode
Printer Mode and language
ANSI/WIN Katakana
Printer Mode and language
ANSI/WIN Russian
Printer Mode
Printer Mode
Printer Mode and language
Nihongo
Printer Mode and language
Russian
sample is on the weighing
pan.
Statistics or Totaling
application.
range
Printer.Settings:
Factory settings
Factory settings
Factory settings
Factory settings
CharSet IBM/DOS Cyrilic
Factory settings
Factory settings
Factory settings
CharSet IBM/DOS Cyrilic
Correction
• Put a new sample or remove
it and put it again on the pan.
• Reset the balance.
• Change the unit back to the
first one
• Sample weight must be in the
range of 70%–130% of the
current average value.
Compatible.Printer.Models
RS-P25, RS-P26, RS-P28
RS-P25
RS-P25
RS-P25