2 6
C e F A R t e n s
7 .4 . 3 s u mm ar y s c r ee n
Summary screen is displayed after last parameter is
adjusted and "oK" button is pressed once. Use "up and
down" to scroll through the different parameters for
visualization only.
7 .4 .4 P a r a met e r s r an g e
Conventional
(C)
Frequency
Preset: 80 Hz
Range: 2–120 Hz
Pulse duration
Preset: 180 µs
Range: 50–300 µs
Time
Preset: – – min.
Range: 1-60, – – min.
P11, P12, P13, P14, P15
Burst
Pulse width/duration
(B)
Frequency
Preset: 2 Hz
Range: 1–5 Hz
Range: 2–120 Hz
Pulse duration 1
Time
Preset: – – min.
Range: 50–300 µs
Range: 1-60, – – min.
Pulse duration 2
Range: 50–300 µs
Modulation time
Preset: – – min.
Range: 1-60, – – min.
Programming
CEFAR TENS
Frequency
modulation
modulation
(PWM)
Frequency
Pulse duration
Preset: 80 Hz
Preset: 180 µs
Range: 50–300 µs
Frequency 1
Preset: 70 µs
Preset: 20Hz
Range: 2–120 Hz
Frequency 2
Preset: 180 µs
Preset: 40Hz
Range: 2–120 Hz
Modulation time
Preset: 3 s
Preset: 5 s
Range: 1–10 s
Range: 5–15 s
Time
Preset: – – min.
Range: 1-60, – – min.
Mixed frequency
(MF)
(FM)
Frequency 1 (Burst)
Preset: 2 Hz
Range: 1–5 Hz
Frequency 2
Preset: 80Hz
Range: 6–120 Hz
Stimulation time
Preset: 3 s
Range: 1–10 s
Time
Preset: – – min.
Range: 1-60, – – min.
Time