Rated maximum output power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watt (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
28 | Mode d'emploi | Elexo Chaise de douche-toilette
Recommended separation distance between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d =1,2 √P
0,12
0,38
1,2
3,8
12
m
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d =1,2 √P
d =2,4 √P
0,12
0,24
0,38
0,76
1,2
2,4
3,8
7,6
12
24