  • FR

Les langues disponibles

5 . C
L E A N I N G
The device and accessories are not disinfected or sterile when delivered.
For cleaning, disinfection or sterilization, the respective Operating Instructions and required cleaning
regulations at the place of use must be observed.
Before cleaning, the device must be switched OFF
and the mains power cable must be disconnected from the mains supply.
Failure to comply may result in electric shock!
If the pump unit is contaminated by failure to use the hydrophobic bacterial filter
or by another cause, the device must be disposed of by the user in accordance
with local regulations.
Do not use cleaning agents and disinfectants based on phenol!
If disregarded, there is a risk of destruction of the plastic materials by fine cracks.
Do not leave the disinfectant in the suction jar.
The plastic material is thereby exposed to unnecessary chemical action
and the strength of the suction jar may be affected.

5.1 Device housing

The surfaces should be cleaned with a damp cloth and a mild detergent.
After cleaning, dry the surfaces with a clean, dry cloth.
For disinfection, only use known agents that do not affect painted surfaces and plastic parts.
The recommendations of the disinfectant manufacturer regarding application and dosage must be
followed.

5.2 Suction jar, lid of suction jar and silicone tubing

-
Autoclave to a maximum of 134° C for a maximum of 15 minutes.
Residues of disinfectants must be removed before autoclaving or neutralized.
Only autoclave dry parts.
Do not stack the parts when autoclaving.
-
Thermal disinfection: do not use alkaline detergents.
-
Chemical disinfection: with aldehyde-based disinfectants, e.g. Glutaraldehyde.
-
Maximum permissible long-term temperature 140° C in steam and 150° C in dry heat.
Part description
Suction jar
Lid with handle
Couplings connectors green and transparent
Couplings green and transparent
Pressure pin
Clamp with threaded pin
Pressure spring
4x O-ring 8.1 x 1.6mm
O-ring silicone 12.42 x 1.78mm
Connection tubing and Patient tubing
Seal for lid of suction jar
Retainer for float of overflow safety device
Float for overflow safety device
Remove any remaining liquid in the float of the overflow safety device
before assembling the lid of suction jar.
Remaining liquid in the vacuum flow may activate the hydrophobic bacterial filter.
Ardo medical AG
Material
PSU Polysulfone
Grivory® GVX 5H
stainless steel
Silicone
EPDM Ethylene-Propylene-Diene (rubber)
POM Polyacetal (Polyoxymethylene)
PP Polypropylene
15
Operating Instructions
Master and Senator
Subject to modifications