Time of application: The models DIALIGHT XL and the
"diagnostic light, white" are designed for intermittent
use. The maximum operation time totals 1 minute. There
shall be 10-minute pauses between the applications.
Improper use of the device/contraindi-
cations for use:
Any other use of use outside the above intended use shall
be considered improper. The manufacturer is not
responsible for damage resulting from such use.
DIALIGHT XL and the „diagnostic lamp, white" can not be
used for eye examination!
Warnings:
• If the diagnostic light is visibly damaged, it should be
removed from clinical use.
• If the light falls, it may break.
• Never immerse diagnostic lights in liquid.
• Avoid direct contact of the diagnostic lights with the
tongue or mucosa.
• The device should not be used in strong magnetic fi elds
(MRI).
• Do not use the device in fl ammable or explosive
environment (for example, oxygen or anesthetics).
• Diagnostic lights, like any electrical diagnostic
devices,are subject to special precautions regarding
electromagnetic compatibility.
• Use only original KaWe accessories and spare parts.
An error may cause extensive heat posing a threat
of burn injury. In case of strong heating, immedia-
tely turn off the device and do not use it again.
Secondary treatment limitations:
No special requirements. The service life of the product is
usually defi ned based on the degree of wear and
use-related damage. Prompt secondary processing of the
tool is recommended after its application.
Transportation and storage:
Keep the tool away from dust, humidity and
contamination during storage and transportation.
Ambient temperature:
Operation: from +10 °C to +35 °C
Storage: -10 °C to +45 °C
Transportation: -20 °C to +50 °C
Relative air humidity:
30 % to 75 %
Cleaning:
Clean the light surface with a damp or wet cloth.
Disinfection:
The light surface can be disinfected with an
alcohol-soaked cloth.
Sterilization:
Sterilization is not possible.
Control and functional check:
Before each use, check the diagnostic light for faultless
operation. Do not use the device if damage is found.
Make sure the moving parts are fi rmly fi xed.
Disposal:
The product should be disposed of at a separate
collection point for electrical appliances and electronics.