RECOMMENDED CLEANING PROCEDURE
Whatever the procedure followed for the decontamination is, the use of a sterile sleeve is compulsory
during all the duration of the surgical operation.
(Steranios 2%)
You must select one of the two solutions indicated for this type of equipment and use it exclusively during
all the lifetime of the product for the pre-disinfection and disinfection processes.
This camera is not autoclavable
This camera is not compatible with automatic washers-disinfectors.
Existing alkaline solution for the pre-disinfection of some medical devices are FORBIDDEN for the pre-
disinfection of our cameras.
It is peremptory to thoroughly rinse with sterile water the parts which have been in contact with the
disinfectant.
Use unweaved compresses for the drying of the distal lens to avoid scratching it.
The procedures described in this chapter are given to you as advice. They can in no case substitute for
official recommendations or directives.
Decontamination being linked to the products, methods and/or selected tools, it remains under the entire
responsibility of the people in charge.