7
Maintenance
8
Troubleshooting
Problem
Pointer does not move
when pump is operated
or drops back again
very quickly
Pointer goes beyond
the 100 % mark or
pump does not return
fully to its stop
Wrong indication
Other malfunction
16
Table 2: Maintenance intervals
When
Condensate trap con-
tains water
At time of tank servic-
ing or tank cleaning
Repair work may only be carried out by qualified, specially trained
personnel.
Table 3: Troubleshooting
Possible reason
Connections are not
airtight or the capillary
lines are damged and
leaking
Filling process
Capillary line is either
blocked or has a kink
in it
Condensate trap full
The wrong measuring
range has been ad-
justed
Measuring range ad-
justed wrongly
Zero correction not
correctly set
Activity
Empty condensate trap
Check Unitel for correct function and,
if necessary, have the instrument re-
adjusted
Remedy
Tighten connections and
check capillary lines for air
tightness
Take measurement after the
tank filling process
Check that capillary line has
no kinks in it and clear any
blockage
Install condensate trap
Empty condensate trap
Check tank dimensions and
correct measuring range, see
chapter 5.2, page 13
See above
Reset Zero correction, see
chapter 5.2, page 13
Return the device to the manu-
facturer