Maintenance And Cleaning; Wear Parts And Accessories - Xylem SI Analytics ScienceLine IDS Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

ScienceLine
Cleaning
After cleaning
T 814254
05/2014

Maintenance and cleaning

During operation, a small amount of reference electrolyte leaks through
the junction from the combination electrode into the test sample ( except
ScienceLine A7780 IDS). If the level of reference electrolyte becomes too low
with time, refi ll it through the refi lling opening. Refi lling is very easy with the
enclosed dropping bottle. Proceed as follows:
Cut off the tip of the dropping bottle at a right angle until the opening in the
tip can be seen
Open the refi lling opening of the combination electrode
Press the tip of the dropping bottle into the refi lling opening while turning it
slightly
Pump the reference electrolyte in the shaft using the dropping bottle
Pull the dropping bottle out of the refi lling opening while turning it slightly if
necessary.
Remove water-soluble contamination by rinsing with deionized water. Other
types of contamination have to be removed as follows while the contact time
with the detergents should be kept as short as possible:
Contamination
Fat and oil
Lime and hydroxide
desposits
Proteins
Note
Hydrofl uoric acid, hot phosphoric acid and strong alkaline solutions destroy the
glass membrane.
Rinse the electrode with deionized water and condition it in reference
electrolyte solution for at least 1 hour. Then recalibrate the electrode.

Wear parts and accessories

Description
Reference electrolyte solution
50 ml to fi ll the watering cap
(KCl 3 mol/l, Ag
+
-free)
Pepsin cleaning solution
Note
Detailed information on our wide range of buffer solutions and more acces-
sories is given in the price list of the Laboratory products catalog.
Maintenance and cleaning | Wear parts and accessories
Cleaning procedure
Rinse with water containing household washing-up liquid
Rinse with citric acid (10 % by weight)
Immerse in pepsin cleaning solution for approx. 1 hour.
Note: Make sure the level of the reference
electrolyte is above that of the cleaning solution.
Model
L3008
L5104
Order no.
285138505
285138295
17

Publicité

Table des Matières
loading

Table des Matières