Switch On Supply Voltage; During Operation; Service Interruptions - Dräger PEX 1000 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Operation
3.2

Switch on supply voltage

After having switched on the PEX 1000, all 14 LED segments including both decimal points
are activated for approx. 5 seconds ("lamp test") to make sure that they are not faulty.
During this period internal test routines start running in the PEX 1000, the output signal of
the PEX 1000 is 1 mA.
Then, the display shows two horizontal lines and the currently measured gas concentration
alternately, changing with approx. 1 Hz. During this period the output signal of the
PEX 1000 is 3.4 mA.
After one minute the PEX 1000 is ready for operation.
If the zero point does not fall below more than 1.25 %LEL or 0.25 %LEL, it shows the
currently measured value and the output signal equals the measured value.
3.3

During operation

Corresponding to the gas concentration the PEX 1000 produces a current between 4 and
20 mA, or
Current
0 mA
1 mA
3.4 mA
3.8 mA to 20.5 mA
4 mA
20 mA
20.5 mA
3.4

Service interruptions

In the event of service interruptions, e.g. when performing maintenance and inspection work,
the PEX 1000 is ready for use again after approx. 10 minutes (sensor warm-up time) when the
system is switched on again.
28
Meaning
Cable is broken or power supply has failed
Failure:
Zero point underrun by more than 5 %LEL or 1 %LEL
Electronic fault
Broken cable or short-circuit in the sensor circuit
Calibration signal
Measuring signal in normal operation
Measuring signal zero point
Measuring signal full-scale value
Exceeded by more than 3 % of the full-scale value
Dräger PEX 1000

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

8324835

Table des Matières