Télécharger Imprimer la page

User Notes - Lovibond Scuba II Mode D'emploi

Masquer les pouces Voir aussi pour Scuba II:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 8

User notes

No.
Meaning and possible causes
01
Insufficient light at ZERO. Poss. cause:
Duct contaminated or light path blocked;
sample strongly coloured?
03
Division by 0. Poss. cause: Duct conta-
minated or light path blocked; sample
strongly coloured?
20
Too much light at detector. Poss. cause:
Cap not applied; measurement in bright
sunlight
21
Too much light at detector. Poss. cause:
Cap not applied; measurement in bright
sunlight
22
LowBat occurred during measurement.
Change battery, clean contacts
36
Measured value too small
38
Measured value too large
Important disposal instructions for
batteries and accumulators
EC Guideline 2006/66/EG requires users to
return all used and worn-out batteries
and accumulators. They must not be disposed of in
normal domestic waste. Because our products include
batteries and accumulators in the delivery package
our advice is as follows :
Used batteries and accumulators are not items of
domestic waste. They must be disposed of in a
proper manner. Your local authority may have a
disposal facility; alternatively you can hand them in
at any shop selling batteries and accumulators. You
can also return them to the company which supplied
them to you; the company is obliged to accept them.
26
GB

Publicité

loading