Precautions regarding Electromagnetic Compatibility (EMC)
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided
in the user's manual and in the present document.
The DMX2 Plus/DMX2 Pro complies with the EMC requirements according to IEC 60601-1-2. Radio transmitting equipment, cellular phones, etc. shall not be used
in the close proximity of the device since this could influence the performance of the device. Particular precaution must be considered during use of strong emis-
sion sources such as High Frequency surgical equipment and similar so that e.g. the HF cables are not routed on or near the device.
If in doubt, please contact a qualified technician or Bien-Air Dental.
WARNING!
The use of accessories, transducers and cables other than those specified, with the exception of transducers and cables sold by Bien-Air as replacements
parts for internal components, may result in increased emissions or decreased immunity of the DMX2 Plus/DMX2 Pro.
Guidance and manufacturer's declaration – electromagnetic emissions
The DMX2 Plus/DMX2 Pro is intended for use in the electromagnetic environment specified below. The customer or the user of the DMX2 Plus/DMX2 Pro should as-
sure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Not applicable
flicker emissions
IEC 61000-3-3
Electromagnetic environment - guidance
The DMX2 Plus/DMX2 Pro uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equipment.
The DMX2 Plus/DMX2 Pro is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.